Skip to content

ENDOTAK RELIANCE® G 0177Recalls & reported problems

Boston Scientific Corporation · Steroid-Eluting Bipolar Pace/Sense and Defibrillation Lead · Class 3 · Implant

All models of ENDOTAK RELIANCE G (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (P910073); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 24, 2025Class IOpen, Classified

    Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

    Z-2325-2025 · Root cause: Component design/selection · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P910073. One submission often covers many models, so check the product and lot details in each record.

  • Z-2326-2025 · Jul 24, 2025 · Open, Classified
    ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296
  • Z-2327-2025 · Jul 24, 2025 · Open, Classified
    RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692,

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page