Skip to content

ENDOTAK RELIANCE G/SG, transvenous defibrillation lead… recallZ-2325-2025 · Boston Scientific Corporation · initiated Jul 24, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

FDA root cause category: Component design/selection

What the company is doing

On July 24, 2025, firm notified physicians via Important Medical Device Information letters. The letter provides information on the management of gradually rising daily subthreshold, low-voltage shock impedance (LVSI) pattern associated with calcification of expanded polytetrafluoroethylene (ePTFE) coated single coil (SC) and dual coil (DC) RELIANCE defibrillation leads manufactured by Boston Scientific Corporation between 2002 to 2021. Letter includes recommendations for routine follow-ups of affected leads, and considerations to take if lead replacement is planned. The firm is not recommending any changes to the scheduled follow-up interval for patients with ePTFE lead models.

Recall details

Recall number
Z-2325-2025
Status
Open, Classified
Initiated by the company
Jul 24, 2025
Posted by the FDA
Aug 20, 2025
Recalling firm
Boston Scientific Corporation
Quantity
247,702 US; 344,670 OUS
Product code
LWS
FDA submissions
P910073
Distribution
Worldwide Distribution. US nationwide. International distribution worldwide.

Product description

ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Model 0161 UDI-DI/GTIN 00802526372766; Model 0164 UDI-DI/GTIN N/A; Model 0165 UDI-DI/GTIN 00802526377259; Model 0167 UDI-DI/GTIN N/A 00802526373251; Model 0170 UDI-DI/GTIN 00802526373282 00802526373299; Model 0171 UDI-DI/GTIN 00802526373374 00802526373381; Model 0172 UDI-DI/GTIN 00802526373466; Model 0173 UDI-DI/GTIN 00802526373558 00802526373565; Model 0174 UDI-DI/GTIN 00802526373640; Model 0175 UDI-DI/GTIN 00802526373732 00802526373749; Model 0176 UDI-DI/GTIN 00802526376726 00802526376733; Model 0177 UDI-DI/GTIN 00802526376818 00802526376825; Model 0180 UDI-DI/GTIN 00802526411977 00802526422027 00802526433597 00802526531149 00802526540943 00802526540950 00802526591518; Model 0181 UDI-DI/GTIN 00802526411984 00802526422034 00802526433689 00802526433726 00802526433757 00802526491009 00802526502903 00802526531156; Model 0182 UDI-DI/GTIN 00802526422041 00802526531163; Model 0183 UDI-DI/GTIN 00802526422058 00802526531170; Model 0184 UDI-DI/GTIN 00802526412011 00802526422065 00802526433986 00802526433993 00802526491313 00802526531187 00802526541261 00802526583438; Model 0185 UDI-DI/GTIN 00802526412028 00802526413438 00802526422072 00802526434082 00802526434099 00802526434105 00802526434150 00802526434174 00802526470622 00802526491405 00802526491412 00802526491429 00802526503306 00802526531194 00802526541346 00802526591563 00802526612541; Model 0186 UDI-DI/GTIN 00802526412035 00802526422089 00802526434198 00802526531200; Model 0187 UDI-DI/GTIN 00802526412042 00802526422096 00802526531217

Devices named in this recall

7 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

41 devices cleared under P910073 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97318).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page