ACUITY Whisper View™ 4647Recalls & reported problems

Boston Scientific Corporation · Guide Wire with Hydrophilic Coating · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00802526512933), received between Sep 26, 2019 and Jun 10, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Difficult to Remove” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Difficult to Remove2
  • Material Integrity Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 10, 2022Difficult to Remove; Material Integrity Problem; Difficult to Remove; Material Integrity Problem

    IT WAS REPORTED THAT THE CATHETER WOULD NOT WITHDRAW WHEN ATTEMPTING TO REMOVE AFTER THE LEFT VENTRICULAR (LV) LEAD PLACEMENT. THE PHYSICIAN THEN PULLED HARDER AND THE DISTAL TIP SHRED OFF REMAINING DISTAL TO LV LEAD. MOREOVER, THE TOP OF THE WIRE REMAINS STATIONARY IN CORONARY SINUS AS CONFIRMED THROUGH FLUOROSCOPY. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14655571 ↗
  • InjurySep 26, 2019Difficult to Remove; Material Integrity Problem; Difficult to Remove; Material Integrity Problem

    IT WAS REPORTED THAT DURING THE LEAD IMPLANT PROCEDURE THE TIP OF THE GUIDEWIRE BROKE OFF WHEN RETRACTED. PHYSICIAN WAS NOT ABLE TO REMOVE THE FOREIGN BODY. NO ADVERSE PATIENT EFFECTS WERE REPORTED. AT THIS TIME, THE PRODUCT HAS NOT BEEN RETURNED. IF THE PRODUCT IS RETURNED, ANALYSIS WILL BE PERFORMED AND THIS REPORT WOULD BE UPDATED AT THAT TIME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9125585 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →