INGEVITY™+ 7842Recalls & reported problems

Boston Scientific Corporation · Pace/Sense Lead · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific Ingevity+ pacing lead is a wire placed in the heart to deliver pacing signals. The reports often describe the lead failing to make the heart respond to pacing, high pacing thresholds, and leads moving out of place. FDA records list 5,794 reports, 21 deaths, 4,151 injuries and 1,622 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
5,995
Deaths reported
21
Injuries reported
4,272
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5,995 adverse event reports that list this exact device (DI 00802526604522), received between Apr 9, 2020 and Sep 30, 2026 — classified as 21 deaths, 4,272 injuries, 1,702 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (1,204 reports).

28 other recalls cover devices cleared under the same FDA submission (P150012); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 89 reportsDec 2023: 77 reportsJan 2024: 103 reportsFeb 2024: 91 reportsMar 2024: 91 reportsApr 2024: 82 reportsMay 2024: 85 reportsJun 2024: 108 reportsJul 2024: 98 reportsAug 2024: 102 reportsSep 2024: 97 reportsOct 2024: 124 reportsNov 2024: 113 reportsDec 2024: 92 reportsJan 2025: 121 reportsFeb 2025: 108 reportsMar 2025: 118 reportsApr 2025: 72 reportsMay 2025: 107 reportsJun 2025: 119 reportsJul 2025: 129 reportsAug 2025: 121 reportsSep 2025: 113 reportsOct 2025: 141 reportsNov 2025: 136 reportsDec 2025: 144 reportsJan 2026: 119 reportsFeb 2026: 107 reportsMar 2026: 151 reportsApr 2026: 126 reportsMay 2026: 123 reportsJun 2026: 125 reportsJul 2026: 98 reportsAug 2026: 121 reportsSep 2026: 168 reportsOct 2026: 0 reports202420252026168
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1,757
  • High Capture Threshold1,204
  • Device Dislodged or Dislocated1,139
  • Failure to Capture1,100
  • Defective Device679
  • Insufficient Device Problem Information553
  • Mechanical Problem431
  • High impedance421

Patient problems

  • No Clinical Signs, Symptoms or Conditions3,643
  • Unspecified Infection1,297
  • Cardiac Perforation243
  • Perforation220
  • No Code Available190
  • Sepsis158
  • Bradycardia124
  • Pain87

Recent adverse event reports

The 25 most recent of 5,995 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Material Integrity Problem

    It was reported that this brady lead was not successfully implanted due to helix got bent after multiple placement attempts. This brady lead was replaced with a non Boston Scientific model. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949522 ↗
  • MalfunctionSep 30, 2026Fracture; High impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947592 ↗
  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this right ventricular (RV) lead was part of a system revision due to infection. There were no additional adverse patient effects reported. The RV lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50947540 ↗
  • InjurySep 30, 2026Signal Artifact/Noise

    It was reported that this right ventricular (RV) lead was surgically abandoned due to concern with the RV channel. Technical Services (TS) reviewed the presenting electrogram (EGM) and found deflections on the RV channel. The noise was not being sensed by the device. A new RV lead was successfully implanted. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946741 ↗
  • InjurySep 30, 2026Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945082 ↗
  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this right ventricular (RV) lead had perforated through the wall of patient's heart. This lead was explanted and a new RV lead was placed. No additional adverse patient effects were reported.

    Patient problems: Perforation

    MAUDE report 50944843 ↗
  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this right ventricular (RV) lead was explanted due to an unknown reason. A new lead was implanted. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943824 ↗
  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50942569 ↗
  • InjurySep 29, 2026Failure to Capture; Device Dislodged or Dislocated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939402 ↗
  • MalfunctionSep 29, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938702 ↗
  • InjurySep 29, 2026Insufficient Device Problem Information

    It was reported right ventricular (RV) lead was surgically abandoned due to an unknown product performance anomaly. A new lead was implanted. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937002 ↗
  • InjurySep 29, 2026Signal Artifact/Noise; Failure to Capture; Fracture; Impedance Problem

    It was reported that this lead exhibited fracture, noisy signals, loss of capture (LOC) and pacing impedance issues. Subsequently, this lead was explanted and successfully replaced. No additional adverse patient effects were reported. This product is expected to be returned for analysis.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934463 ↗
  • MalfunctionSep 29, 2026Material Integrity Problem; Positioning Problem

    It was reported that this Brady lead was not successfully implanted due to the septum that was chosen during implant was too thick and the numbers were getting worse after the catheter was slit. Furthermore, the lead was pulled out, and it was noted that the lead was stretched. A new lead was implanted. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933204 ↗
  • MalfunctionSep 29, 2026Signal Artifact/Noise; High impedance; Pacing Problem

    It was reported that this right ventricular (RV) lead exhibited noisy signals due to unexpected programming. Additionally, a red alert was found for high out of range pace impedance measurements. Technical Services (TS) was consulted and found the system to be operating in unipolar rather than bipolar with no previous history of a lead safety switch. TS believed a bigger lead anomaly was present but have no confirmation at this time. TS noted that pacing inhibition greater than 2 seconds had occurred at times. No additional adverse patient effects were reported. This RV lead remains in service.

    Patient problems: Asystole

    MAUDE report 50932547 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50930528 ↗
  • MalfunctionSep 29, 2026Insufficient Device Problem Information

    It was reported that this right ventricular (RV) lead was attempted due to unknown product performance reason. This RV lead was replaced and not implanted. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927891 ↗
  • MalfunctionSep 29, 2026Nonstandard Device; Device Contamination with Body Fluid

    It was reported that this Brady lead was not implanted successfully due to tissue stuck in helix and a new Brady lead of the same model was placed. This lead will be returned by the hospital. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927811 ↗
  • InjurySep 28, 2026Failure to Capture

    Patient problems: Bradycardia

    MAUDE report 50923790 ↗
  • InjurySep 28, 2026Mechanical Problem; Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923406 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left bundle branch (LBB) lead was part of a system revision due to infection. There were no additional adverse patient effects reported. The LBB lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50917391 ↗
  • MalfunctionSep 28, 2026Failure to Capture

    t was reported that the cardiac resynchronization therapy pacemaker (CRT-P) and this left ventricular (LV) lead suspended a left ventricular automatic threshold (LVAT) test. Technical Services (TS) reviewed device data and confirmed the LVAT was suspended due to fusion beats. Also, the presenting electrogram (EGM) exhibited loss of capture on the LV lead due to the LV and right ventricular (RV) beats falling on top of each other. TS recommended further evaluation in clinic. This LV lead remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915460 ↗
  • MalfunctionSep 28, 2026Insufficient Device Problem Information

    It was reported that this Brady lead was not implanted successfully due to product performance anomaly and a new Brady lead of the same model was placed. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911013 ↗
  • MalfunctionSep 25, 2026Failure to Read Input Signal; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906244 ↗
  • MalfunctionSep 25, 2026Insufficient Device Problem Information

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50904683 ↗
  • MalfunctionSep 25, 2026Failure to Capture; High impedance; Failure to Read Input Signal

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50904251 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P150012. One submission often covers many models, so check the product and lot details in each record.

  • Z-1770-2026 · Mar 19, 2026 · Open, Classified
    Boston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy) labeled as: 1. ESSENTIO SR SL, Model Number L100; 2. ESSENTIO DR SL
  • Z-0095-2026 · Aug 20, 2025 · Open, Classified
    Model Number L111, ESSENTIO DR SL MRI Pacemaker
  • Z-0084-2026 · Aug 20, 2025 · Open, Classified
    ACCOLADE DR EL MRI (Model L331)
  • Z-0094-2026 · Aug 20, 2025 · Open, Classified
    Model Number L110, ESSENTIO SR SL MRI Pacemaker
  • Z-0082-2026 · Aug 20, 2025 · Open, Classified
    ACCOLADE DR SL MRI (Model L311)
  • Z-0091-2026 · Aug 20, 2025 · Open, Classified
    Model Number L231, PROPONENT DR EL MRI Pacemaker
  • Z-0088-2026 · Aug 20, 2025 · Open, Classified
    Model Number L210 PROPONENT SR SL Pacemaker
  • Z-0097-2026 · Aug 20, 2025 · Open, Classified
    Model Number L131, ESSENTIO DR EL MRI Pacemaker
  • Z-0089-2026 · Aug 20, 2025 · Open, Classified
    Model Number L211 PROPONENT DR SL MRI Pacemaker
  • Z-0081-2026 · Aug 20, 2025 · Open, Classified
    ACCOLADE SR SL MRI (Model L310)
  • Z-0824-2025 · Dec 12, 2024 · Open, Classified
    ESSENTIO DR Pacemaker, SL (Model Number L101, SL MRI (Model Number L111), EL (Model Number L121), and EL MRI (Model Number L131)
  • Z-0823-2025 · Dec 12, 2024 · Open, Classified
    PROPONENT DR Pacemaker, SL (Model Number L201), (VDD) SL (Model Number L209), SL MRI (Model Number L211), EL (Model Number L221), and EL MRI (Model Number L231).
  • Z-0822-2025 · Dec 12, 2024 · Open, Classified
    ACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Number L331)
  • Z-1999-2021 · Jun 3, 2021 · Open, Classified
    Proponent pacemaker, Models: a) L200 b) L201 c) L209 d) L210 e) L211 f) L221 g) L231
  • Z-1998-2021 · Jun 3, 2021 · Open, Classified
    Essentio pacemaker, Models: a) L100 b) L101 c) L110 d) L111 e) L121 f) L131

…and 13 more.

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →