Tendril™ STS (REF 2088TC-58)Recalls & reported problems
St. Jude Medical, Inc. · Pacing Lead · Class 3 · Implant
All models of Tendril (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This St. Jude Tendril STS pacing lead has 12,645 adverse event reports, including 204 deaths and 6,327 injuries. Many describe the lead over-sensing heart signals, failing to pace the heart, or picking up noise, and some describe broken or dislodged leads. Some patients report infections, discomfort or dizziness, and several reports describe leads being replaced. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 12,659
- Deaths reported
- 199
- Injuries reported
- 6,335
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12,659 adverse event reports that list this exact device (DI 05414734502894), received between Jun 11, 2018 and Sep 30, 2026 — classified as 199 deaths, 6,335 injuries, 6,125 malfunctions.
The most frequently reported specific device problem is “Over-Sensing” (3,350 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Over-Sensing3,350
- Failure to Capture1,829
- Signal Artifact/Noise1,773
- Adverse Event Without Identified Device or Use Problem1,665
- Break1,479
- Device Dislodged or Dislocated1,381
- Capturing Problem1,295
- High impedance1,258
Patient problems
- No Clinical Signs, Symptoms or Conditions7,570
- No Known Impact Or Consequence To Patient2,085
- Unspecified Infection1,233
- No Consequences Or Impact To Patient356
- Discomfort307
- Dizziness262
- Arrhythmia232
- Insufficient Information211
Recent adverse event reports
The 25 most recent of 12,659 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Difficult to Remove
New information received notes that the patient presented to the hospital for a generator changeout procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50955162 ↗MalfunctionSep 30, 2026High impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50955147 ↗MalfunctionSep 30, 2026Over-Sensing
Patient problems: Dizziness
MAUDE report 50953561 ↗MalfunctionSep 30, 2026Over-Sensing
During an in-clinic follow up, sensing noise was observed on the right ventricular (RV) Lead. The patient is pacer dependent. No intervention has been performed at this time. The patient is stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947200 ↗MalfunctionSep 30, 2026Break; Retraction Problem
It was reported that the patient presented for an implant procedure. During the implant procedure the physician needed to reposition the lead, but the helix was unable to be retracted when fixated in the myocardium. The lead was pulled from the heart, and it was noted that the helix was deformed and broken. The lead was replaced with a new lead and procedure was completed successfully. Patient was discharged following the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946866 ↗InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient presented to the hospital due to infection at the device site. The pacemaker, right atrial lead and right ventricular lead were explanted and replaced. The patient was in stable condition throughout the procedure.
Patient problems: Unspecified Infection
MAUDE report 50942775 ↗InjurySep 30, 2026Signal Artifact/Noise; Low impedance
It was reported that the patient presented to the hospital for a right ventricular (RV) lead extraction procedure due to noise and low pacing impedance measurements. The RV lead was explanted and replaced. The patient was stable throughout.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942609 ↗MalfunctionSep 29, 2026High impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941401 ↗InjurySep 28, 2026Failure to Sense; Capturing Problem; Device Dislodged or Dislocated
It was reported that the patient presented for a follow-up in clinic. Interrogation revealed that both the right ventricular (RV) and right atrial (RA) lead exhibited loss of sensing and were capturing inappropriately. Chest X-ray further confirmed that both the RV and RA leads were completely dislodged and wrapped underneath the pacemaker as the patient was a Twiddler. Both RV and RA leads were explanted and replaced. The patient was asymptomatic and remained stable throughout the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926693 ↗MalfunctionSep 28, 2026Difficult to Fold, Unfold or Collapse
It was reported that during a patient procedure the lead's helix retracted and could not be unscrewed. The lead was not used, and a new lead was implanted satisfactorily. The patient was stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917626 ↗InjurySep 28, 2026Low impedance
It was reported that the patient presented to clinic for follow up, the right ventricle (RV) lead exhibited low pacing impedance. The RV lead was capped and replaced on (b)(6)2026. The patient was discharged with no reports of adverse consequences.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910951 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient presented with an infection at the device site. No erosion was observed. The pacemaker, right atrial lead and right ventricular lead were explanted. The patient remained stable throughout the procedure.
Patient problems: Unspecified Infection
MAUDE report 50910896 ↗MalfunctionSep 28, 2026Over-Sensing
It was reported that the Right Ventricular (RV) lead exhibited oversensing of suspected electromagnetic interference (EMI) noise. No intervention was reported. There were no patient symptoms reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910760 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient's pacemaker, right atrial (RA) lead and right ventricular (RV) lead were explanted on (b)(6) 2026 due to infection. Temporary pacemaker was placed until the infection was resolved; the patient was subsequently implanted with a new pacemaker system on (b)(6) 2026. The patient was in stable condition.
Patient problems: Unspecified Infection
MAUDE report 50910672 ↗MalfunctionSep 28, 2026Difficult to Fold, Unfold or Collapse
It was reported that the patient presented for an implant procedure. During the procedure, the helix of the lead failed to extend. As a result, the lead was not implanted and was replaced during the same procedure. The patient remained stable throughout the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910649 ↗InjurySep 25, 2026Insufficient Device Problem Information
It was reported that the patient's pacemaker and the right ventricular (RV) lead were explanted due to unknown reason. A replacement RV lead was implanted. The patient was stable throughout the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50897906 ↗InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient presented to the clinic with redness and swelling at the implant pocket site secondary to an infection. The physician elected to explant the entire system, including the pacemaker, right ventricular (RV) lead, and right atrial (RA) lead, on (b)(6) 2026. The patient remained in stable condition.
Patient problems: Unspecified Infection
MAUDE report 50893665 ↗InjurySep 24, 2026Fracture; High impedance
It was reported that high pacing impedance was noted on the right ventricular (RV) lead. Lead fracture was observed. The RV lead was capped and replaced on (b)(6) 2026. There were no adverse events to the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50884018 ↗InjurySep 24, 2026High impedance; Capturing Problem
It was reported that the patient presented to the clinic for a scheduled follow-up. Upon interrogation, the patient's right ventricular (RV) lead exhibited a high capture threshold and high pacing impedance. The RV lead was explanted and replaced on (b)(6) 2026. The patient remained in stable condition.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50878350 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported that patient presented to clinic with fever and chills. The patient experienced an adverse event of infection. The reported infection was determined not to be associated with any Abbott product or procedure. Blood cultures were obtained and positive for Staphylococcus bacteremia. Due to the infection, the pacemaker (PM), the right ventricle (RV) and atrial leads were explanted. The cause of the infection was attributed to Staphylococcus bacteremia.
Patient problems: Fever; Unspecified Infection
MAUDE report 50877501 ↗MalfunctionSep 24, 2026Failure to Capture
It was reported that during an implant procedure, the lead exhibited elevated pacing thresholds and failed to achieve capture despite programming at high pacing outputs. Lead repositioning was attempted; however, no improvement in electrical performance was observed. The physician elected to not use the lead, and a new lead was implanted. The patient was stable throughout the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50877321 ↗MalfunctionSep 24, 2026Break; Device Dislodged or Dislocated
It was reported that patient presented to operating room for lead implant. During procedure when the operator was splitting the peel-away sheath the lead was pulled out and the helix was damaged. The lad was not used and was replaced with a new lead. The patient was stable following the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50876950 ↗MalfunctionSep 24, 2026Break; Low impedance
It was reported that the patient presented for an implant procedure. During final parameter testing, it was observed that the right ventricular (RV) lead exhibited low pacing impedance and an insulation break. The RV lead was not implanted and was replaced with an alternative RV lead. The procedure was completed successfully. The patient remained in stable condition throughout the event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50876721 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient was scheduled for valve surgery due to bacteremia, aortic valve (AO) endocarditis, and mitral valve (MV) endocarditis. The patient presented the day prior to the scheduled valve surgery for explantation of the implanted pacemaker system, including the right ventricular (RV) lead and right atrial (RA) lead. During the extraction procedure, transesophageal echocardiography (TEE) revealed perforation of the superior vena cava (SVC) and a progressive pericardial effusion. Pericardiocentesis was performed to drain the fluid; however, the attempt was unsuccessful. Cardiothoracic (CT) surgery subsequently proceeded with an emergency pericardial window, and the SVC perforation was surgically repaired. The entire pacemaker system was successfully explanted. The patient remained stable.
Patient problems: Cardiac Perforation; Low Blood Pressure/ Hypotension; Pericardial Effusion; Unspecified Infection
MAUDE report 50876637 ↗MalfunctionSep 23, 2026Capturing Problem; Device Sensing Problem
It was reported that the patient presented for an implant procedure. During the procedure, the right ventricle lead exhibited unsatisfactory pacing threshold and R-wave sensing values. The lead was not used and replaced. The patient was in stable condition.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50868170 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P960013 on Jun 20, 1997. Submitted by Abbott Medical as “Tendril Dx Model 1388t & 1388k Endocardial Pacing Leads”. The FDA review took 406 days. The FDA has since approved 130 supplements (changes to the device or its labeling), most recently on Oct 9, 2026. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
The FDA lists this type of device as life-sustaining or life-supporting.
FDA definition of this device type: Drug eluting permanent RV or RA pacemaker electrodes are used with compatible pacemakers or implantable cardioverter defibrillators (ICD) or cardiac rhythm therapy (CRT) devices to deliver pacing pulses to the heart.
Common questions about the Tendril™ STS
Has the Tendril™ STS been recalled?
No FDA recall record in our data names this exact device identifier (DI 05414734502894), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Tendril™ STS?
The FDA's MAUDE database lists 12,659 adverse event reports for this device, received between Jun 11, 2018 and Sep 30, 2026. In those reports, the most common specific device problems are “Over-Sensing” (3,350), “Failure to Capture” (1,829) and “Signal Artifact/Noise” (1,773); the most common patient problems are “Unspecified Infection” (1,233), “Discomfort” (307) and “Dizziness” (262). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Tendril™ STS?
199 reports are classified as deaths, 6,335 as injuries and 6,125 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Tendril™ STS FDA approved?
It was approved through premarket approval (PMA) P960013 on Jun 20, 1997.
Is the Tendril™ STS MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Tendril™ STS, and is it still sold?
The FDA's device database lists St. Jude Medical, Inc. as the company responsible for it (brand Tendril™, model STS). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes436 deaths19,593 reports
- INGEVITY™ MRI 7741Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes36 deaths7,885 reports
- INGEVITY™+ 7841Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes15 deaths7,442 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes190 deaths6,771 reports
- INGEVITY™+ 7842Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes21 deaths6,021 reports
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes40 deaths4,992 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes50 deaths4,987 reports
- INGEVITY™ MRI 7742Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes24 deaths3,970 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.