The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00803622103056), received between Jan 17, 2018 and Nov 15, 2023 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Disconnection” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K972503); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Disconnection2
  • Material Twisted/Bent2
  • Detachment of Device or Device Component1
  • Increase in Pressure1
  • Material Deformation1
  • Material Too Rigid or Stiff1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Patient Involvement2
  • Insufficient Information1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 15, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18139449 ↗
  • MalfunctionMay 9, 2023Increase in Pressure; Material Twisted/Bent; Increase in Pressure; Material Twisted/Bent

    LIVA NOVA USA INC HAS RECEIVED A REPORT THAT, DURING PROCEDURE, WHEN INJECTING CARDIOPLEGIA SOLUTION IN THE PATIENT'S AORTIC ROOT, THE PERFUSIONIST DETECTED UNUSUAL HIGH PRESSURE ABOVE 600 MMHG. INJECTION WAS STOPPED. A SECOND CARDIOPLEGIA INJECTION ATTEMPT CONFIRMED THE HYPERTENSION. MEDICAL TEAM ELECTED TO CHANGE OUT THE CARDIOPLEGIA CANNULA. THE PROCEDURE WAS COMPLETED WITH NO ISSUE. AFTER REMOVAL OF THE CANNULA, A WATER INJECTION INTO THE CANNULA SHOWED A REFLUX ON THE LATERAL LINE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16893618 ↗
  • MalfunctionJun 22, 2021Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    SEE INTIAL REPORT.

    Patient problems: Insufficient Information

    MAUDE report 12041599 ↗
  • MalfunctionSep 18, 2020Disconnection; Disconnection

    LIVANOVA USA INC HAS BEEN INFORMED THAT, DURING PRIMING, THE LUERLOCK CONNECTION HAS DETACHED FROM THE SKEWER. THERE IS NOT REPORT OF ANY PATIENT INJURY.

    Patient problems: No Patient Involvement

    MAUDE report 10554121 ↗
  • MalfunctionAug 27, 2020Disconnection; Disconnection

    LIVANOVA USA INC HAS BEEN INFORMED THAT, DURING A PROCEDURE, THE LUERLOCK CONNECTION HAS DETACHED FROM THE SKEWER. THERE IS NOT REPORT OF ANY PATIENT INJURY.

    Patient problems: No Patient Involvement

    MAUDE report 10458332 ↗
  • InjuryJan 17, 2018Material Too Rigid or Stiff; Material Too Rigid or Stiff

    LIVANOVA RECEIVED A REPORT THAT THE AORTIC ROOT CANNULA CAUSED TRAUMA TO THE PATIENT'S AORTA. DURING FURTHER COMMUNICATION, IT WAS REVEALED THAT THE USER FEELS THE MATERIAL IS HARD AND THE AORTA IS "CRUSHED" WHEN THE CANNULA IS ATTACHED.

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 7198694 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K972503. One submission often covers many models, so check the product and lot details in each record.

  • Z-2757-2016 · Aug 12, 2016 · Terminated
    Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices