Cardioplegia solution administration adaptor CA-xxxxx (UDI 00803622103650)Recalls & reported problems

Livanova USA, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00803622103650), received between Oct 11, 2016 and Oct 11, 2016 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Increase in Pressure2
  • No Flow2

Patient problems

  • Blood Loss2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 11, 2016Fluid/Blood Leak; Increase in Pressure; No Flow; Fluid/Blood Leak; Increase in Pressure; No Flow

    SORIN GROUP RECEIVED A REPORT THAT THE CARDIOPLEGIA ADAPTER CANNULA WAS BLOCKED AND CAUSED OVER-PRESSURIZATION AND BLOOD LEAKAGE DURING A PROCEDURE. THE DEVICE WAS SWITCHED OUT TWICE, AND THE ISSUE RECURRED BOTH TIMES. ON THE THIRD OCCURRENCE, THE USER NOTICED THAT THE Y-CONNECTOR WAS BLOCKED. THE PATIENT LOST A SIGNIFICANT AMOUNT OF BLOOD, AND THE PROCEDURE WAS EXTENDED BY 40 MINUTES.

    Patient problems: Blood Loss

    MAUDE report 6018809 ↗
  • MalfunctionOct 11, 2016Fluid/Blood Leak; Increase in Pressure; No Flow; Fluid/Blood Leak; Increase in Pressure; No Flow

    SORIN GROUP RECEIVED A REPORT THAT THE CARDIOPLEGIA ADAPTER CANNULA WAS BLOCKED AND CAUSED OVER-PRESSURIZATION AND BLOOD LEAKAGE DURING A PROCEDURE. THE DEVICE WAS SWITCHED OUT TWICE, AND THE ISSUE RECURRED BOTH TIMES. ON THE THIRD OCCURRENCE, THE USER NOTICED THAT THE Y-CONNECTOR WAS BLOCKED. THE PATIENT LOST A SIGNIFICANT AMOUNT OF BLOOD, AND THE PROCEDURE WAS EXTENDED BY 40 MINUTES.

    Patient problems: Blood Loss

    MAUDE report 6018807 ↗

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices

All CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices →