Skip to content

FlexFlow Venous Cannula 200-200Recalls & reported problems

Livanova USA, Inc. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00803622125812), received between Oct 3, 2017 and Oct 3, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Issue” (1 report).

1 other recall cover devices cleared under the same FDA submission (K052081); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Issue1
  • Device Operates Differently Than Expected1
  • Material Deformation1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 3, 2017Device Issue; Device Operates Differently Than Expected; Material Deformation; Device Issue

    LIVANOVA RECEIVED A REPORT THAT THE TIPS OF SEVERAL FLEX FLOW VENOUS CANNULAE WERE FOUND TO BE SHARP/POINTED WHEN THEY SHOULD BE BLUNT. THIS ISSUE WAS NOTED DURING SETUP AND THE PRODUCT WAS NOT USED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 6910181 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K052081. One submission often covers many models, so check the product and lot details in each record.

  • Z-0076-2018 · Sep 28, 2017 · Terminated
    FlexFlow Venous Cannula Model/Catalog Number 200-200 The FlexFlow Venous Cannula is a single use sterile device. It is a size 23 French with an outside diameter of 0.30 inch (7.7 m

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices

All CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page