Skip to content

DuraTip® DURA30-USRecalls & reported problems

Parkell, Inc. · 30K DuraTip® Universal Slim Insert · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00810000522927), received between Oct 28, 2025 and Apr 6, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Handling Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Handling Problem2
  • Material Fragmentation2
  • No Apparent Adverse Event2
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 6, 2026Material Fragmentation; No Apparent Adverse Event; Device Handling Problem

    PARKELL, INC. BECAME AWARE OF AN INCIDENT INVOLVING A PARKELL INSERT, THE DURA30-US, ON MARCH 9, 2026. THE INSERT WAS IN USE BY A STUDENT WHEN IT BROKE OFF IN A PATIENT'S MOUTH. IT WAS CAUGHT IN THE TRAP AND IMMEDIATELY SUCTIONED OUT. NO INJURY WAS INITIALLY REPORTED. NO TREATMENT WAS SOUGHT. IMAGES OF THE INSERT WERE PROVIDED FOR EVALUATION. THE INSERT WAS RETURNED WITHOUT THE TIP. THE INSERT IS PART OF THE STUDENT INSTRUMENT PACKAGE. AS PART OF THE RESOLUTION, THE HYGIENE COORDINATOR AT THE UNIVERSITY WAS INSTRUCTED TO REMIND CLINICAL STUDENTS TO ADHERE TO THE PERFORMANCE GUIDE TO ENSURE PROPER DEVICE HANDLING AND USAGE PARAMETERS, MINIMIZING THE RISK OF MECHANICAL FATIGUE OR BREAKAGE. WE ARE FILLING THIS REPORT IN AN ABUNDANCE OF CAUTION TO COMPLY WITH 21 CFR 803.10(C). WE WILL FILE A FOLLOW-UP REPORT IF WE BECOME AWARE OF ANY ADDITIONAL RELEVANT INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24794385 ↗
  • MalfunctionDec 16, 2025Fracture; Fracture

    ON (B)(6) 2025, WE RECEIVED A COMPLAINT INVOLVING OUR DURA30-US INSERT. THIS INSERT IS AN ACCESSORY TO OUR "TURBO" ULTRASONIC SCALER. INITIAL COMPLAINT INDICATED THAT THE TIP BROKE. SUBSEQUENTLY WE DISCOVERED THAT, UPON INITIATION OF PROCEDURE, THE INSERT TIP BROKE IN THE PATIENT'S MOUTH. THERE WAS NO INJURY TO EITHER THE HYGIENIST OR THE PATIENT. THE TIP WAS IMMEDIATELY RECOVERED. THE INSERT WAS RETURNED FOR EVALUATION MINUS THE BROKEN TIP. THE INSERT IS UNDER EVALUATION. WE ARE FILING THIS REPORT WITH AN OVERABUNDANCE OF CAUTION TO COMPLY WITH 21 CFR 803.10(C). A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23815789 ↗
  • MalfunctionOct 28, 2025Material Fragmentation; No Apparent Adverse Event; Device Handling Problem; Material Fragmentation; No Apparent Adverse Event; Device Handling Problem

    ON OCTOBER 10TH, 2025, PARKELL BECAME AWARE OF AN INCIDENT INVOLVING OUR INSERT, DURA30-US - DURATIP® 30K UNIVERSAL SLIM INSERT. THIS INSERT IS AN ACCESSORY TO OUR TURBO ULTRASONIC SCALER. THE INSERT WAS IN USE BY A DENTAL STUDENT WHEN IT BROKE OFF IN THE PATIENT'S MOUTH. NO INJURY WAS INITIALLY REPORTED. NO TREATMENT WAS SOUGHT. THE UNIVERSITY IS KEEPING THE INSERT AND WILL NOT RETURN IT FOR EVALUATION. WE HAVE RECEIVED IMAGES OF THE INSERT BOTH ALONE AND IN OUR PERFORMANCE GUIDELINE TOOL. EVALUATING THESE IMAGES, IT IS OUR EXPERT OPINION THAT THE INSERT WAS USED BEYOND PERFORMANCE GUIDELINE REQUIREMENT OF 3 MM OF WEAR. IFU INSTRUCTION FOR PERFORMANCE GUIDELINE - IF THE END OF THE TIP DOES NOT EXTEND FULLY ACROSS THE 3 MM INDICATOR GROOVE (FIGURE 4) OR IS FLUSH WITH THE LEFT SIDE OF THE GROOVE, OR IS SHORT OF THE GROOVE, THE TIP HAS REACHED 3 MM OF WEAR AND SHOULD BE REMOVED FROM SERVICE AND REPLACED. WE ARE FILLING THIS REPORT IN AN ABUNDANCE OF CAUTION TO COMPLY WITH 21 CFR 803.10(C). WE WILL FILE A FOLLOW-UP REPORT IF WE BECOME AWARE OF ANY ADDITIONAL RELEVANT INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23401745 ↗

Other Scaler, Ultrasonic devices

All Scaler, Ultrasonic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page