Impulse Dynamics (usa) Inc. · OPTIMIZER Lite Implantable Pulse Generator for the Treatment of Heart Failure System consists of: (1) OPTIMIZER Lite Implantable Pulse Generator (IPG) (1) Allen #2 torque wrench (1) Port Plug · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00810003380142), received between Dec 3, 2024 and Sep 17, 2026 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Misassembly by Users” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 5, 2024Open, Classified

    OPTIMIZER devices may cease to deliver CCM therapy if the device incorrectly detects a charging error. This may cause patients to experience heart failure symptoms similar to before implantation of the device.

    Z-1173-2024 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Misassembly by Users2
  • Installation-Related Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Bacteremia1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Misassembly by Users

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50811010 ↗
  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    An Impulse Dynamics field representative reported that after a second lead dislodgement occurred in a patient implanted with an OPTIMIZER Lite IPG, the patient and doctor agreed to have the IPG and leads removed. The rep attended the explant procedure. The explanted IPG was returned to Impulse Dynamics USA in Marlton, NJ for evaluation. There is no evidence at this time that the IPG was malfunctioning at the time of explant. It is unclear as to the reason(s) why the leads repeatedly dislodged in this patient (ex. anatomy, Twiddler's, etc.) prior to system explant.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50793418 ↗
  • InjuryAug 14, 2026Misassembly by Users

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50258200 ↗
  • InjuryNov 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2025, AN IMPULSE DYNAMICS FIELD REPRESENTATIVE WAS INFORMED THAT A PATIENT HAD THEIR OSM IPG AND LEADS EXPLANTED THAT DAY. THE PATIENT HAD DEVELOPED AN INFECTION AROUND THEIR IMPLANTED PORT-A-CATH DEVICE, WHICH HAD BEEN IMPLANTED DUE TO THE PATIENT'S ONGOING CANCER TREATMENT. THE PHYSICIAN ELECTED TO EXPLANT ALL OF THE PATIENT'S IMPLANTED DEVICES, WHICH INCLUDED AN OSM IPG. HOSPITAL BLOOD CULTURES IDENTIFIED BACTEREMIA AS THE TYPE OF INFECTION AROUND THE PORT-A-CATH. THE EXPLANTED OSM IPG AND LEADS ARE CURRENTLY BEING HELD UNDER QUARANTINE AT THE HOSPITAL. THE IPG IS EXPECTED TO BE RECEIVED BY IMPULSE DYNAMICS USA IN (B)(6) IN THE COMING WEEKS TO BE FURTHER EVALUATED. THERE IS NO EVIDENCE AT THIS TIME TO INDICATE THE IPG CAUSED OR CONTRIBUTED TO THE PATIENT'S INFECTION.

    Patient problems: Bacteremia

    MAUDE report 23513489 ↗
  • MalfunctionDec 3, 2024Installation-Related Problem; Installation-Related Problem

    DURING THE IMPLANTATION OF AN OPTIMIZER LITE (OL) IMPLANTABLE PULSE GENERATOR, IT WAS DETERMINED THAT THE PATIENT'S LEADS REQUIRED INTRAOPERATIVE ADJUSTMENT. THE SURGEON RETRACTED THE SET SCREWS OF THE V1 AND V2 LEADS, BUT UPON RE-ENGAGEMENT AND TIGHTENING, THE V1 LEAD WAS UNABLE TO BE INSERTED. THIS WAS ATTRIBUTED TO THE INABILITY OF ITS DISTAL SET SCREW TO BE TIGHTENED. THE SURGEON STATED "THE SET SCREW HAD BEEN RETRACTED TOO FAR AND HAD BECOME MISALIGNED WITH THE BORE AND THREADS." AT THAT POINT, THE DEVICE WAS DISCONNECTED FROM BOTH LEADS AND SWAPPED WITHOUT INCIDENT. THE UNUSED DEVICE WILL BE RETURNED TO (B)(4) FOR FURTHER EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20832380 ↗