Skip to content

OPTIMIZER recallZ-1173-2024 · Impulse Dynamics (USA) Inc · initiated Jan 5, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Impulse Dynamics (USA) Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

OPTIMIZER devices may cease to deliver CCM therapy if the device incorrectly detects a charging error. This may cause patients to experience heart failure symptoms similar to before implantation of the device.

FDA root cause category: Software design

What the company is doing

On January 2, 2024, the firm sent letters to field representatives to be given to affected physicians. Summary of the safety communication can be found online: https://impulse-dynamics.com/safety-information/ The firm recommends all patients with implant card showing model CCM X11, who have experienced the "A9" charge error, charge their implanted device to no more than 75% of full capacity to circumvent the issue. See link above for instructions. If a patient encounters an A9 error code, CCM therapy has been suspended and must be manually reset. Patients should schedule time with an Impulse Dynamics representative or their physician to manually reset their device. The firm is working on a software update to address this issue. Patients and clinicians may determine if a specific device is affected by providing the serial number of any specific device to the Technical Support Hotline at 866-312-5370.

Recall details

Recall number
Z-1173-2024
Status
Open, Classified
Initiated by the company
Jan 5, 2024
Posted by the FDA
Feb 22, 2024
Recalling firm
Impulse Dynamics (USA) Inc
Quantity
1,469 total units
Product code
QFV
FDA submissions
P180036S021, P180036S025, P180036S007
Distribution
Domestic US distribution nationwide. International distribution to Austria Croatia Czech Republic Germany Italy Macau Poland Romania Russia Spain Sweden.

Product description

OPTIMIZER model CCM X11 implantable pulse generator (IPG) devices - Smart Mini and Lite

Code information (models, lots, serial numbers)

OPTIMIZER Smart Mini UDI-DI: 00810003380098 OPTIMIZER Lite UDI-DI: 00810003380142 All units potentially affected. The available software versions that allow the issue to manifest are 1.5.0 and 1.9.1.

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page