Skip to content

Paradise® Catheter PRDS-064-02Recalls & reported problems

Recor Medical, Inc. · PRDS-064-02 Paradise® Catheter 4.2mm · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810008950005), received between Dec 15, 2025 and Dec 15, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P220023); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Thrombosis/Thrombus1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 15, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    EVENT: RENAL ARTERY THROMBOSIS EVENT DESCRIPTION: A FEMALE PATIENT UNDERWENT ULTRASOUND RENAL DENERVATION PROCEDURE ON (B)(6) 2025. DURING THE PROCEDURE, MULTIPLE ATTEMPTS WERE MADE TO ACCESS THE CHALLENGING, TORTUOUS LEFT RENAL ARTERY USING THREE DIFFERENT PARADISE BALLOONS, WHICH REPEATEDLY DISLODGED DUE TO RESISTANCE CAUSED BY VESSEL TORTUOSITY. PARADISE CATHETER WAS NEVER INFLATED NOR WAS ANY SONICATION PROVIDED THE LEFT RENAL ARTERY. ALTHOUGH THE RETURNED DEVICES HAD TEARS WHICH APPEARED TO BE DUE TO TORTUOUS ANATOMY, THIS IS HIGHLY UNLIKELY RELATED TO THE ADVERSE EVENT. AS PER PHYSICIAN, THE ANTICOAGULATION DOSE WAS INADEQUATE. POST PROCEDURAL RENAL ANGIOGRAM NOTED A CLOT PARTIALLY OCCLUDING THE LEFT RENAL ARTERY. THE CLOT WAS TREATED WITH UNILATERAL THROMBECTOMY USING PENUMBRA PUMP (CAT6) WHICH RESTORED THE BLOOD FLOW. FOLLOWING THE THROMBECTOMY PROCEDURE, THE PATIENT WAS TRANSFERRED TO THE ICU. A CONTRAST-ENHANCED CT SCAN WAS PERFORMED NOTED AREAS OF INFARCTION IN THE LEFT KIDNEY. PATIENT CONDITION IMPROVED AND DISCHARGED UNDER STABLE CONDITION 3 DAYS LATER. SUBOPTIMAL ANTICOAGULATION DOSE ALONG WITH FORCED INSERTION OF PARADISE CATHETERS AGAINST CHALLENGING ANATOMY, ATHERO

    Patient problems: Thrombosis/Thrombus

    MAUDE report 23812610 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P220023. One submission often covers many models, so check the product and lot details in each record.

  • Z-1655-2026 · Jan 29, 2026 · Open, Classified
    Brand Name: Paradise Ultrasound Renal Denervation System Product Name: Paradise Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter rena
  • Z-1654-2026 · Jan 29, 2026 · Open, Classified
    Brand Name: Paradise Ultrasound Renal Denervation System Product Name: Paradise Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter rena

Other Ablation Catheter, Renal Denervation devices

All Ablation Catheter, Renal Denervation devices →

Sources and updates