Paradise® Catheter PRDS-066-02Recalls & reported problems
Recor Medical, Inc. · PRDS-066-02 Paradise® Catheter 6mm · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810008950029), received between Dec 22, 2025 and Dec 22, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Expiration Date Error” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (P220023); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Expiration Date Error2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2025Expiration Date Error; Expiration Date Error
EVENT: EXPIRED DEVICE USED. EVENT DESCRIPTION: A PATIENT UNDERWENT ULTRASOUND RENAL DENERVATION PROCEDURE ON (B)(6) 2025 USING A PARADISE CATHETER THAT HAD EXPIRED ON 29 NOV 2025. THE PARADISE CATHETER WAS INSPECTED PRIOR TO USE AND THE PACKAGING AND SEAL WAS INTACT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND NO PROCEDURAL COMPLICATION OR PATIENT INJURY/HARM REPORTED FOR THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23873903 ↗MalfunctionDec 22, 2025Expiration Date Error; Expiration Date Error
A PATIENT UNDERWENT ULTRASOUND RENAL DENERVATION PROCEDURE ON (B)(6) 2025 USING A PARADISE CATHETER THAT HAD EXPIRED ON (B)(6) 2025. THE PARADISE CATHETER WAS INSPECTED PRIOR TO USE AND THE PACKAGING AND SEAL WAS INTACT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND NO PROCEDURAL COMPLICATION OR PATIENT INJURY/HARM REPORTED FOR THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23873792 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P220023. One submission often covers many models, so check the product and lot details in each record.
- Z-1654-2026 · Jan 29, 2026 · Open, ClassifiedBrand Name: Paradise Ultrasound Renal Denervation System Product Name: Paradise Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter rena
- Z-1655-2026 · Jan 29, 2026 · Open, ClassifiedBrand Name: Paradise Ultrasound Renal Denervation System Product Name: Paradise Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter rena
Other Ablation Catheter, Renal Denervation devices
- Paradise® Catheter PRDS-068-02Recor Medical, Inc. · Ablation Catheter, Renal Denervation2 recalls0 reports
- Symplicity Spyral™ RDN016Medtronic, Inc. · Ablation catheter, renal denervation5 reports
- Paradise® Catheter PRDS-065-02Recor Medical, Inc. · Ablation Catheter, Renal Denervation4 reports
- Paradise® Catheter PRDS-067-02Recor Medical, Inc. · Ablation Catheter, Renal Denervation2 reports
- Paradise® Catheter PRDS-064-02Recor Medical, Inc. · Ablation Catheter, Renal Denervation1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page