Skip to content

IMPEDE Embolization Plug IMP-07Recalls & reported problems

Shape Memory Medical Inc. · IMPEDE Embolization Plug, 7mm · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00810041650016), received between Jan 4, 2022 and Feb 11, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Insufficient Device Problem Information1
  • Migration1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Clinical Signs, Symptoms or Conditions1
  • Peripheral Edema1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2026Migration

    DURING AN ARTERIOVENOUS (AV) SHUNT CASE ON (B)(6) 2025 IN TAIWAN, THE PHYSICIAN IMPLANTED AN IMP-07 DEVICE NEAR THE WRIST OF THE PATIENT. WHEN THE PATIENT RETURNED FOR A FOLLOW-UP ON (B)(6) 2026 DUE TO CONCERNS OF SWELLING, IT WAS FOUND THE DEVICE HAD MIGRATED PROXIMALLY (UP TOWARDS THE ELBOW AREA) WHICH OCCLUDED THE MAIN VESSEL. THE PHYSICIAN PERFORMED A CUT-DOWN TO REMOVE THE IMP-07 DEVICE AND IT REMAINS EXPLANTED FROM THE PATIENT. NO SERIOUS CLINICAL EVENT OCCURRED TO THE PATIENT AND THE PATIENT IS STABLE.

    Patient problems: Peripheral Edema

    MAUDE report 24318996 ↗
  • MalfunctionJan 20, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE REPORTED DATE OF THE EVENT IS (B)(6) 2021, ALTHOUGH THE INITIAL PROCEDURE TOOK PLACE ON (B)(6) 2021. ON (B)(6) 2021, A 2-DEBRANCH TEVAR WAS BEING PERFORMED AND AN IMP-10 AND IMP-07 WAS USED TO EMBOLIZE THE LEFT SUBCLAVIAN ARTERY (LSCA) AT THE SAME TIME. THE IMP-10 WAS INTRODUCED INTO THE BODY FIRST, FOLLOWED BY THE IMP-07 IN PARALLEL; BOTH DEVICES WERE REPORTED TO BE DELIVERED INTO THE TARGET SITE. THE OPERATION WAS COMPLETED, ALTHOUGH IT WAS NOT CONFIRMED AT THE TIME THAT EMBOLIZATION OF THE LSCA WAS ACHIEVED WITH THE IMP-10 DEVICE. ON (B)(6) 2021, DURING THE POST-OPERATIVE EXAMINATION USING CONTRAST-ENHANCED COMPUTED TOMOGRAPHY, A TYPE II ENDOLEAK WAS FOUND AT THE LSCA. ON (B)(6) 2021, AN OPERATION WAS PERFORMED TO TREAT THE TYPE II ENDOLEAK FOUND DURING THE POST-OPERATIVE EXAMINATION USING ADDITIONAL NON-SMM EMBOLIC COIL DEVICES (3 INTERLOCK COILS FROM BOSTON SCIENTIFIC). THESE DEVICES WERE DEPLOYED TO THE LSCA FROM THE VICINITY OF THE IMPEDE EMBOLIZATION PLUGS, USING AN IMPRESS 5FR DIAGNOSTIC CATHETER FROM MERIT MEDICAL. IT WAS CONFIRMED THAT THE INTERLOCK COILS WERE DELIVERED, COMPLETE EMBOLIZATION WAS ACHIEVED, THE TYPE II ENDOLEAK HAD DISAPPEARED, AND THE OPERATION WAS C

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13311458 ↗
  • MalfunctionJan 4, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT UNDERWENT BYPASS SURGERY TO TREAT A POPLITEAL ANEURYSM. DUE TO THE PRESENCE OF A PATENT POPLITEAL VESSEL (7-8 MM DIAMETER), THE PHYSICIAN ATTEMPTED EMBOLIZATION VIA DELIVERY OF AN IMP-07 DEVICE. AFTER OBSERVING NO VESSEL OCCLUSION, AN IMP-FX-08 DEVICE WAS DELIVERED TO THE SAME TARGET AREA. TEN MINUTES AFTER THE IMP-FX DELIVERY, RESIDUAL FLOW WAS STILL OBSERVED WHICH RESULTED IN SEVERAL COOK COILS BEING USED TO COMPLETE THE CASE. NO ADVERSE EVENT TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13160195 ↗

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page