Skip to content

EdgeTaper Platinum SETF421HTRecalls & reported problems

US Endodontics, LLC · Sterile Endodontic Rotary Files · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 2, 2025Open, Classified

    Due to devices being irradiated above the established specification which my result in package integrity issues.

    Z-1114-2025 · Root cause: Under Investigation by firm · FDA record ↗

Other FILE, PULP CANAL, ENDODONTIC devices

All FILE, PULP CANAL, ENDODONTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page