Skip to content

GalaFLEX 3D 100042-06Recalls & reported problems

Tepha, Inc. · GalaFLEX 3D Scaffold, 7.5 cm x 21.0 cm, Large Oval (Contents: 2) · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00810054980315), received between Jun 10, 2022 and Jul 31, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K161092); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient-Device Incompatibility3
  • Device Appears to Trigger Rejection1

Patient problems

  • Bacterial Infection1
  • Foreign Body Reaction1
  • Impaired Healing1
  • Seroma1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 31, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    AS REPORTED, THE PATIENT UNDERWENT A BREAST IMPLANT EXCHANGE ON (B)(6) 2025. ABOUT 5 WEEKS POSTOP, THE PATIENT PRESENTED IN WITH AN UNSPECIFIED INFECTION VERY LATE IN THE HEALING PROCESS ON THE RIGHT BREAST. ON (B)(6) 2025, PATIENT UNDERWENT REMOVAL OF THE MESH AS IT WAS POTENTIAL AFFECTED BY THE INFECTION AND WAS TREATED WITH ANTIBIOTICS.

    Patient problems: Unspecified Infection

    MAUDE report 22676247 ↗
  • InjuryMar 19, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AS REPORTED, PATIENT UNDERWENT BILATERAL BREAST AUGMENTATION WITH IMPLANT OF GALAFLEX 3D MESH ON (B)(6) 2025. AT THE 4 WEEK FOLLOW UP APPOINTMENT THE PATIENT PRESENTED WITH EXPOSED IMPLANT ON THE RIGHT BREAST POST IMPLANT. IT WAS REPORTED THAT IT DID NOT HEAL, AND NO ISSUE WAS FOUND ON THE LEFT BREAST POST IMPLANT. SURGEON PLANNED TO REMOVE THE IMPLANT ON (B)(6) 2025 AND TO REPLACE WITH NEW IMPLANT ONCE THE HEALED. NO FURTHER INFORMATION HAS BEEN PROVIDED ON WHETHER THE IMPLANT WAS REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 21644280 ↗
  • InjuryNov 8, 2024Patient-Device Incompatibility; Patient-Device Incompatibility

    AS REPORTED, ON (B)(6) 2024 PATIENT UNDERWENT A BILATERAL BREAST AUGMENTATION WITH GALAFLEX 3D MESH IMPLANTED. IT WAS REPORTED THAT POST IMPLANT, PATIENT WAS PRESENTED WITH SYMPTOMS (FEVER AND POCKETING) OF REJECTING EITHER THE IMPLANTS OR THE GALAFLEX 3D THEREBY UNDERWENT REMOVAL OF GALAFLEX 3D MESH ON (B)(6) 2024. AS REPORTED, THE PATIENT WAS PRESCRIBED ANTIBIOTICS POST OP.

    Patient problems: Bacterial Infection; Foreign Body Reaction

    MAUDE report 20640623 ↗
  • InjuryJun 10, 2022Patient-Device Incompatibility; Patient-Device Incompatibility

    PER CLINICAL TRIAL CP-(B)(4): ON (B)(6) 2016, PATIENT UNDERWENT BILATERAL REVISION MASTOPEXY WITH AUTO-AUGMENTATION USING AN 'OTHER' INCISION : AREOLA, WISE PATTERN SUPERIOR PEDICLE. 2 PIECES OF GALASHAPE (GALAFLEX 3D) WERE USED. CONCOMITANT PROCEDURES WERE LIPOSUCTION, MASTOPEXY, AUTO-AUGMENTATION. ON (B)(6) 2016, BILATERAL BREAST SEROMA WAS REPORTED. SAE. PATIENT HAD PROCEDURES OF ASPIRATION OF SEROMA BEGINNING (B)(6) 2016. SEROSANGUINOUS DRAINAGE EVACUATED FROM BOTH BREASTS ON SEVERAL DATES. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2017. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED SEROMA IS LIKELY RELATED TO THE DEVICE AND SERIOUS.

    Patient problems: Seroma

    MAUDE report 14654250 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K161092. One submission often covers many models, so check the product and lot details in each record.

  • Z-0948-2017 · Nov 21, 2016 · Terminated
    GalaSHAPE 3D; product code SH3D03, a medium size oval GalaSHAPE" 3D is indicated for use as a bioresorbable scaffold for soft tissue support and to repair, elevate and reinforce de

Other Surgical Film devices

All Surgical Film devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page