Skip to content

GalaFLEX LITE 100027-10Recalls & reported problems

Tepha, Inc. · GalaFLEX LITE Scaffold, 25.0 cm x 30.0 cm, Rectangle · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810054980827), received between Apr 2, 2024 and Apr 2, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Patient-Device Incompatibility1

Patient problems

  • Bacterial Infection1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 2, 2024Patient-Device Incompatibility; Patient-Device Incompatibility

    AS REPORTED, THE PATIENT UNDERWENT BILATERAL IMPLANT EXCHANGE AND MASTOPEXY PROCEDURE WITH THE IMPLANT OF TWO GALAFLEX LITE MESHES IN LEFT AND RIGHT BREASTS ON (B)(6) 2023. IT WAS REPORTED THAT THE PATIENT WAS DIAGNOSED WITH STAPHYLOCOCCUS INFECTION IN THE LEFT BREAST AND UNDERWENT REVISION SURGERY BY REMOVING THE GALAFLEX LITE MESHES FROM BOTH BREASTS ON (B)(6) 2023. THE SURGEON DOES NOT KNOW WHAT COULD HAVE CAUSED THE PATIENT'S INFECTION. IT WAS REPORTED THAT PATIENT WAS TREATED WITH IV ANTIBIOTICS AND THE INFECTION WAS CLEARED.

    Patient problems: Bacterial Infection

    MAUDE report 19024496 ↗
  • InjuryApr 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, THE PATIENT UNDERWENT BILATERAL IMPLANT EXCHANGE AND MASTOPEXY PROCEDURE WITH THE IMPLANT OF TWO GALAFLEX LITE MESHES IN LEFT AND RIGHT BREASTS ON (B)(6) 2023. IT WAS REPORTED THAT THE PATIENT WAS DIAGNOSED WITH STAPHYLOCOCCUS INFECTION IN THE LEFT BREAST AND UNDERWENT REVISION SURGERY BY REMOVING THE GALAFLEX LITE MESHES FROM BOTH THE LEFT AND RIGHT BREASTS ON (B)(6) 2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19021454 ↗

FDA clearance and classification

The FDA regulates product code OOD (surgical film) under 21 CFR 878.3300 as a Class II device. 21 CFR 878.3300 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: To be implanted to reinforce soft tissue or bone where weakness exists. Intended to be used wherever temporary wound support is required, to reinforce soft tissues where weakness exists in the urological, gynecological, or gastrointestinal anatomy, or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result. The absorbable protective film also may help minimize the potential for tissue attachment to the device in case of direct contact with the viscera.

Common questions about the GalaFLEX LITE 100027-10

Has the GalaFLEX LITE 100027-10 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00810054980827), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the GalaFLEX LITE 100027-10?

The FDA's MAUDE database lists 2 adverse event reports for this device, received on Apr 2, 2024. In those reports, the most common specific device problem is “Patient-Device Incompatibility” (1); the most common patient problem is “Bacterial Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the GalaFLEX LITE 100027-10?

None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the GalaFLEX LITE 100027-10 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.

Who makes the GalaFLEX LITE 100027-10, and is it still sold?

The FDA's device database lists Tepha, Inc. as the company responsible for it (brand GalaFLEX LITE, model 100027-10). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Surgical Film devices

All Surgical Film devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page