Skip to content

SOL-M 10ml Luer Lock Syringe Sterile Convenience Tray P180010TURecalls & reported problems

Sol-Millennium Medical, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00810062434305), received between May 16, 2025 and Jun 19, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Contamination” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination3
  • Crack1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2026Contamination

    CUSTOMER REPORTED THAT DURING THE 100% VISUAL INSPECTION PROCESS, SEVERAL UNITS WERE REJECTED FOR "PARTICULATE/MATERIAL IN THE OTHER CATEGORY, SUSPENDED IN SOLUTION OR ADHERED TO INTERNAL SURFACE OF CONTAINER". THIS IMPACTED FOUR COMPOUNDED LOTS. THE PARTICULATE(S) WERE ANALYZED BY THE QC LAB AND DETERMINED TO BE A POLYESTER FIBER/NATURAL FIBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25561754 ↗
  • MalfunctionMay 5, 2026Contamination

    CUSTOMER REPORTED THAT THE FOREIGN MATERIAL FOUND IN SYRINGE TRAYS. CONTAMINATION IDENTIFICATION ANALYSIS PROVIDED BY THE CUSTOMER INDICATED THAT THE OBSERVED FOREIGN MATERIAL CONSISTED OF: · FIBROUS PARTICLES CONSISTENT WITH POLYESTER FIBERS. · LUBRICANT-TYPE PARTICLES CONSISTENT WITH SILICONE-BASED MATERIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25089793 ↗
  • MalfunctionMay 16, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    CUSTOMER REPORTED THE SYRINGES EXPERIENCED CRACKING DURING USE, WHICH LED TO PRODUCT LEAKAGE AND A LOSS OF PRIMARY PACKAGE INTEGRITY. PRODUCT USED WITH ROPIVACAINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035724 ↗
  • MalfunctionMay 16, 2025Contamination; Contamination

    CUSTOMER REPORTED: WE OBSERVED RUST-LIKE PARTICLES IN ONE OF THE 10ML SYRINGES FROM LOT# 04301002. PLEASE FIND THE ATTACHED IMAGE AND A SMALL VIDEO. PLEASE LET US KNOW THE WAY FORWARD TO ADDRESS THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035642 ↗