Prestige Coil System PRES0204CXPPLTRecalls & reported problems
Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant
All models of Prestige Coil System (97)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810068567342), received between Mar 16, 2023 and May 22, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Migration” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K220699); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 8, 2025Class IIOpen, Classified
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
Z-1989-2025 · Root cause: Vendor change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Migration2
- Failure to Advance1
- Premature Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2026Failure to Advance; Migration
IT WAS REPORTED THAT: "THE PHYSICIAN WAS DOING A DISTAL SPLENIC EMBOLIZATION AT AROUND 1AM. SHE MENTIONS THAT THE ANATOMY TO GET TO THE BLEED WAS QUITE TORTUOUS. SHE WAS USING A 2.0FR PROGREAT MICROCATHETER. SHE HAD ALREADY USED SEVERAL PRESTIGE COILS UP TO THIS POINT WITH NO ISSUE. THE CURRENT PRODUCT SKU INCIDENT HAS NOTHING TO DO WITH THE COIL IMPLANT ITSELF BUT THE PUSHER. SHE DEPLOYED THIS 2MM X 4CM PRESTIGE PLUS COIL IN THE LAST VESSEL AND WHEN SHE WENT TO REMOVE THE PUSHER DELIVERY SYSTEM FROM THE MICROCATHETER SHE MET A LITTLE RESISTANCE FOLLOWED BY NO RESISTANCE AT ALL, THINKING THAT EVERYTHING WAS OK. SHE TRIED TO FLUSH THE MICROCATHETER WITH SALINE, SHE NOTICED THAT THERE WAS HEAVY RESISTANCE TRYING TO FLUSH IT, THINKING THAT MAYBE THE COIL MAY NOT HAVE BEEN FULLY DEPLOYED. SHE THEN LOOKED DID A LIVE FLUORO SHOT (NO CONTRAST) TO CONFIRM THAT THE COIL WAS FULLY REMOVED. SHE TRIED A SECOND TIME TO FLUSH THE MICROCATHETER WITH SALINE, USING A BIT MORE FORCE AND FELT THE MICROCATHETER CLEAR OUT. THE PHYSICIAN THEN DECIDED TO DO ONE MORE POST EMBOLIZATION CONTRAST RUN AND NOTICED WHAT LOOKED TO BE THE DISTAL PART OF THE PUSHER DELIVERY SYSTEM. SHE NOTES THAT THE RADIOPAQUE MA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25262096 ↗MalfunctionMar 16, 2023Migration; Premature Separation; Migration; Premature Separation
IT WAS REPORTED THAT: "THE FIRST COIL, 2/4 REF #PRES0204CXPPLT LOT# F220701319. DETACHED FROM THE INTRODUCER WITHOUT THE DETACHMENT DEVICE AND CAUSED NON TARGET EMBO IN THE DISTAL RIGHT HEPATIC ARTERY AND COULDN'T BE RETRIEVED." UPDATE 03MAR2023: "THE FINAL OUTCOME OF THE CASE WAS SUCCESSFUL THERE WAS NO HARM TO THE PATIENT. THE MIGRATED COIL ACTUALLY SHUTDOWN THE VESSEL AND SHE WAS ABLE TO TREAT THE TARGET ZONE WITH PARTICLES. THE FIRST COIL THAT DETACHED WAS BECAUSE OF THE PACK AND IT KNOTTED UPON ITSELF. SHE THINKS WHEN SHE WENT TO REPOSITION THE COIL SHE DID NOT MOVE MC TO TAKE OUT THAT KNOT OUT AND THEREFORE THE COIL DETACHED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16554026 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K220699. One submission often covers many models, so check the product and lot details in each record.
- Z-1042-2024 · Jan 3, 2024 · Open, ClassifiedThe Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to em
Other Device, Vascular, For Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page