Skip to content

Prestige Coil System PRES3530CXPPLTRecalls & reported problems

Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant

All models of Prestige Coil System (97)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810068567458), received between Sep 20, 2023 and Nov 20, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K220699); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 8, 2025Class IIOpen, Classified

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    Z-1989-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Premature Separation1
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2024Premature Separation; Premature Separation

    IT WAS REPORTED THAT: "THE PHYSICIAN WAS ADVANCING THE THIRD COIL OF THE CASE AND IT BECAME STUCK IN THE MICRO-CATHETER WHEN TRYING TO PULL BACK THE COIL PUSHER BROKE IN THE VESSEL. THE PHYSICIAN THEN USED SALINE TO FORCE THE REST OF THE COIL IMPLANT INTO THE VESSEL BUT THE PUSHER BROKEN PART WAS NOT ABLE TO BE DISLODGED." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20735789 ↗
  • MalfunctionSep 20, 2023Separation Problem; Separation Problem

    IT WAS REPORTED THAT: "DEVICE WAS BEING UTILIZED AND WOULD NOT DEPLOY. DEVICE WAS THEN REMOVED BUT COIL BROKE WHILE BEING REMOVED GI BLEED NO LOCATION GIVEN, PRODUCT NEVER GOT OUT OF SHEATH. PRODUCT WAS TESTED BEFORE INTRODUCED INTO PATIENT." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED. "BROKE INSIDE OF THE PATIENT AND DOCTOR WAS ABLE TO PUSH INTO PLACE WAS DURING A GI BLEED, COILS WAS NOT CAUSE OF THE BLEED".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17786268 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K220699. One submission often covers many models, so check the product and lot details in each record.

  • Z-1042-2024 · Jan 3, 2024 · Open, Classified
    The Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to em

Other Device, Vascular, For Promoting Embolization devices