Prestige Coil System PRES0410CPKPLTRecalls & reported problems
Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant
All models of Prestige Coil System (97)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810068567465), received between Nov 1, 2024 and Jul 23, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Premature Separation” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K220699); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Premature Separation2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 23, 2025Premature Separation; Premature Separation
IT WAS REPORTED THAT: "THE PHYSICIAN WAS PERFORMING AN EMBOLIZATION FOR AN ADBDOMEN HEMORRAGE AND WAS USING A PROGREAT 2.4F MICROCATHETER. SHE WENT TO REPOSITION THE CATHETER AND THE COIL BROKE WITHOUT MUCH TENSION. SHE HAD TO LOSE ACCESS SINCE THE COIL WAS COMPROMISED. THE COIL CAN BE RETURNED FOR INSPECTION." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED. UPDATE - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER ON 23JUN2025: "1. DID THE PROCEDURE INVOLVE TORTUOUS ANATOMY? MILD TORTUOUSITY. 2. WILL THE MICROCATHETER BE AVAILABLE FOR RETURN? NO . 3. WHERE DID THE IMPLANT PREMATURELY DETACH? (INSIDE THE AN, INSIDE THE MC? ETC) THE COIL WAS IN THE PATIENT AND HALFWAY STILL IN THE MICROCATHETER. 4. WERE ANY ATTEMPTS MADE TO DETACH THE IMPLANT COIL USING THE DETACHMENT CONTROLLER? NO. 5. CAN YOU CONFIRM IF THE DEVICE WAS IMPLANTED? DEVICE WAS REMOVED FROM PATIENT".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22583429 ↗MalfunctionNov 1, 2024Premature Separation; Premature Separation
IT WAS REPORTED THAT: "THE PHYSICIAN DECIDED TO USE A TRUSELECT 2.0F MICROCATHETER TO COIL THE OUTFLOW OF A RENAL ANEURYSM. SHE SELECTED A 4X10 PRESTIGE PLUS COIL FROM THE JACKSON CONSIGNMENT. THE SHEATH CAME OFF THE COIL AND SHE BEGAN ADVANCING WHEN SHE FELT RESISTANCE, SHE PUSHED THROUGH SLIGHTLY AND THE COIL SEEMED TO LOOP AS IF THE FRONT END WAS STUCK AND THE BACK END WAS STILL ATTACHED TO THE PUSHER COIL. THE COIL THEN BROKE OFF BUT THE TREAD PART OF THE 4X10 STAYED LODGED IN THE MICROCATHETER. THIS LED TO PRE DETACHMENT AND NEED FOR COMPLETELY NEW ACCESS. AFTER REPLACING THE MICROCATHETER, SHE SNARED THE REST OF THE COIL OUT OF THE VESSEL AND PROCEEDED TO USE PENUMBRA (8 COILS) FOR THE REMAINDER OF THE CASE." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20588447 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K220699. One submission often covers many models, so check the product and lot details in each record.
- Z-1989-2025 · May 8, 2025 · Open, ClassifiedPrestige Coil System REFs/UDI-DI codes: PRES0153CXPPLT 00810068567311 PRES1045CPKPLT 00810068567663 PRES1245CPKPLT 00810068567694 PRES1445CPKPLT 00810068567724 PRES0103CXPPLT 00810
- Z-1042-2024 · Jan 3, 2024 · Open, ClassifiedThe Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to em
Other Device, Vascular, For Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page