Skip to content

Prestige Coil System PRES0420CPKPLTRecalls & reported problems

Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant

All models of Prestige Coil System (97)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810068567489), received between Dec 11, 2024 and Dec 11, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Migration” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K220699); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 8, 2025Class IIOpen, Classified

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    Z-1989-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Migration1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2024Migration; Premature Separation; Migration; Premature Separation

    IT WAS REPORTED THAT: "I WAS NOT IN THIS PROCEDURE AS THIS WAS PERFORMED LATE FRIDAY EVENING. I RECEIVED A PHONE CALL FROM THE PHYSICIAN IN THE MIDDLE OF THIS PROCEDURE. HE WAS PERFORMING A GDA EMBOLIZATION. HE WAS USING A 2.4FR PROGREAT MC AND HAD DEPLOYED A FEW OTHER PRESTIGE COILS IN THIS PROCEDURE PRIOR TO THIS INCIDENT. HE NOTED THAT HE DID NOT EXPERIENCE ANY HEAVY RESISTANCE AND THAT THE ANATOMY WAS NOT TORTUOUS HE NOTD THAT AS HE WAS PULLING BACK THE COIL TO ADJUST, IT HAD PRE-DETACHED.THE DISTAL PART OF THE COIL WAS IN THE GDA AND THE MAJORITY OF THE COIL WAS IN THE MC. HE THEN TRIED TO PUSH IT OUT, UNFORTUNATELY IT DID NOT PACK WELL AND THE PROXIMAL END OF THE COIL STARTED TO MOVE TO THE PROPER HEPATIC ARTERY AND INTO A BRANCH OFF OF THE PROPER HEPATIC." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20903593 ↗

FDA clearance and classification

Cleared through Special 510(k) K220699 on Apr 8, 2022 (decision: Substantially Equivalent). Submitted by Bait USA, LLC as “Prestige Coil System”. The FDA review took 29 days. FDA record ↗

The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K220699. One submission often covers many models, so check the product and lot details in each record.

  • Z-1042-2024 · Jan 3, 2024 · Open, Classified
    The Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to em

Common questions about the Prestige Coil System PRES0420CPKPLT

Has the Prestige Coil System PRES0420CPKPLT been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1989-2025 (Class II), began on May 8, 2025: Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Prestige Coil System PRES0420CPKPLT?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 11, 2024. That report lists device problems “Migration” (1) and “Premature Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Prestige Coil System PRES0420CPKPLT?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Prestige Coil System PRES0420CPKPLT FDA approved?

It was cleared through Special 510(k) K220699 on Apr 8, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Prestige Coil System PRES0420CPKPLT MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Prestige Coil System PRES0420CPKPLT, and is it still sold?

The FDA's device database lists Balt USA LLC as the company responsible for it (brand Prestige Coil System, model PRES0420CPKPLT). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page