Prestige Coil System PRES0865CPKPLTRecalls & reported problems
Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant
All models of Prestige Coil System (97)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810068567632), received between Jun 12, 2024 and May 9, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Migration” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K220699); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration2
- Premature Separation2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 9, 2025Migration; Premature Separation; Migration; Premature Separation
IT WAS REPORTED THAT: "PRES0865CPKPLT - IN A GONADAL CASE WITH DR (B)(6) ON (B)(6) 2025 WE USED SEVERAL PRESTIGE COILS. THE FINAL COIL THAT WE USED DEPLOYED EARLY IN THE 2.4 PROGREAT MICROCATHETER. IT SOUNDS LIKE DR (B)(6) AND THE IR TECH FELT SOME RESISTANCE DURING THE DEPLOYMENT. 70% OF THE COIL DEPLOYED CORRECTLY IN THE VESSEL BUT THE LAST 30% DEPLOYED EARLY AND DIDN'T COIL PROPERLY, THE END RESULT BEING A TAIL FLOATING IN THE RENAL VEIN ABOVE THE REST OF THE COILS IN THE GONADAL VEIN. DR (B)(6) TRIED TO PUSH THE TAIL OF THE COIL WITH A WIRE AND WITH A BALLOON BUT WAS UNABLE TO SUFFICIENTLY MOVE THE COIL. SHE FELT THAT THE COIL WAS SECURE ENOUGH AND WOULDN'T TRAVEL BECAUSE THE INITIAL 70% OF THAT COIL CORRECTLY DEPLOYED AND WAS SECURE." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED. UPDATE 11APR2025- ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "1. ANY UPDATES ON THE PATIENT? - I REACHED OUT TO DR (B)(6) TODAY AND SHE SAID THE PATIENT WAS FINE WHEN THEY WENT HOME TUESDAY AND SHE HASN'T HEARD ANY NEWS FROM THE PATIENT SINCE THEN."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21998762 ↗MalfunctionJun 12, 2024Migration; Premature Separation; Migration; Premature Separation
IT WAS REPORTED THAT: "WHILE ADVANCING THE SECOND COIL, AN 8X65, THROUGH A 2.8 PROGREAT THE FELLOW FELT RESISTANCE. THE STAFF DOCTOR SCRUBBED IN TO HELP AND TRIED TO ADVANCE THE COIL, THEN PULL BACK, AND THE COIL DETACHED IN THE MICROCATHETER. ENOUGH OF THE COIL WAS OUT OF THE CATHETER THAT THE PHYSICIAN WAS ABLE TO FLUSH THE REST OUT OF THE CATHETER INTO THE VESSEL WITH A SALINE SYRINGE." UPDATE 15MAY2024 - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "NO TORTUOUS ANATOMY, AND THE PROCEDURE WAS FINISHED WITH PENUMBRA COILS." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19519359 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K220699. One submission often covers many models, so check the product and lot details in each record.
- Z-1989-2025 · May 8, 2025 · Open, ClassifiedPrestige Coil System REFs/UDI-DI codes: PRES0153CXPPLT 00810068567311 PRES1045CPKPLT 00810068567663 PRES1245CPKPLT 00810068567694 PRES1445CPKPLT 00810068567724 PRES0103CXPPLT 00810
- Z-1042-2024 · Jan 3, 2024 · Open, ClassifiedThe Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to em
Other Device, Vascular, For Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.