Prestige Coil System PRES1865CPKPLTRecalls & reported problems
Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant
All models of Prestige Coil System (97)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810068567793), received between Mar 20, 2024 and Sep 11, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Migration” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K220699); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration2
- Premature Separation1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 11, 2024Stretched; Migration; Stretched; Migration
IT WAS REPORTED THAT: "WE ARE STILL FINDING OUT INFORMATION ABOUT THE CASE BUT HERE IS WHAT IS KNOWN: THE PHYSICIAN WAS USING A BSC RENEGADE MICROCATHTER AND A PRESTIGE PLUS COIL WAS DEPLOYED AND UPON REMOVAL OF THE PUSHER WIRE A STRING OF COIL CAME BACK WITH THE PUSHER. HE SPENT 2-3 HOURS SNARING THE BROKEN COIL FROM THE PATIENT WITH NO HARM. UPDATE 17SEP2024 - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "2ND CASE PORTAL VEIN EMBOLIZATION WITH EXTREME TORTUOSITY. MICROCATHETER USED WAS A BOSTON SCIENTIFIC STC DAMAGED COIL PRES1865CPKPLT LOT#F230900859. HE ADVANCED THE COIL INTO THE ABNORMAL VENOUS SHUNT THAT WAS MEASURING AROUND 15MM CROSS SECTIONAL. HE STATED THAT NOT ALL OF THE COIL MADE IT INTO THE SAC AS HE FELT RESISTANCE. WE DID NOT DETERMINE HOW MUCH OF THE 65CM MADE IT INTO THE SAC BUT HE DECIDED TO DETACH THE COIL AND THOUGHT HE WOULD SQUIRT THE REST OUT. HE HAD PROBLEMS REMOVING THE PUSHER ALMOST LIKE IT WAS "STUCK" THEREFORE HE DECIDED TO REMOVE THE PUSHER WITH THE MICROCATHETER. AS BACK OUT THE MICROCATHETER THE COIL ELONGATED AND FOLLOWED EVENTUALLY SQUIRTING OUT AND STAYING IN THE VESSEL HOWEVER NOT IN THE TARGET ZONE. WAS THE CATHETER WAS OUT HE REMOVED THE PUS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20196273 ↗MalfunctionMar 20, 2024Migration; Premature Separation; Migration; Premature Separation
IT WAS REPORTED THAT: "WHEN PLACING THE 7TH COIL IN A SPLENIC ARTERY ANEURYSM (18X65), THE COIL BACKED OUT OF THE ANEURYSM AND BEGAN COILING IN THE INFLOW VESSEL. THE PHYSICIAN CONTINUED TO ADVANCE THE COIL INTO THE VESSEL, BUT THEN DECIDED TO PULL THE COIL BACK INTO THE MICROCATHETER (PROGREAT ALPHA 2.8->2.0) AND TRY TO REGAIN ACCESS INTO THE ANEURYSM TO FINISH PACKING IT. WHEN THE PHYSICIAN WAS RETRACTING THE COIL, HE ANNOUNCED THAT IT WAS NO LONGER ATTACHED. HE THEN TRIED TO SNARE THE COIL AND IT BROKE, DISTAL TO THE CATHETER IN THE SPLENTIC. HE THEN BEGAN TO PULL THE WINDING OF THE COIL. WE TRIED FOR HOURS TO SNARE." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18942032 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K220699. One submission often covers many models, so check the product and lot details in each record.
- Z-1989-2025 · May 8, 2025 · Open, ClassifiedPrestige Coil System REFs/UDI-DI codes: PRES0153CXPPLT 00810068567311 PRES1045CPKPLT 00810068567663 PRES1245CPKPLT 00810068567694 PRES1445CPKPLT 00810068567724 PRES0103CXPPLT 00810
- Z-1042-2024 · Jan 3, 2024 · Open, ClassifiedThe Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to em
Other Device, Vascular, For Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page