Skip to content

Prestige Coil System PRES0750HPKPLRecalls & reported problems

Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant

All models of Prestige Coil System (97)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810068567861), received between Dec 27, 2024 and Dec 27, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Premature Separation” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K220699); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 8, 2025Class IIOpen, Classified

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    Z-1989-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 27, 2024Premature Separation; Premature Separation

    IT WAS REPORTED THAT: "PRES0750HPKPL WAS DEPLOYED INTO A SPLENIC ARTERY AND WOULD NOT DETACH. 10 PLUS ATTEMPTS WERE MADE AND SECOND DETACHMENT HANDLE TO NO AVAIL. THE COIL WAS RETRACTED. DURING THE COIL BEING REMOVED IT PREMATURELY DETACHED AND SNAPPED HALF WAY DOWN THE IMPLANT. THE REST OF THE COIL IN THE MICROCATHETER WAS ATTEMPTED TO BE PUSHED IN WITH A COIL PUSHER AND WAS ULTIMATELY FLUSHED IN TO FINISH THE CASE." UPDATE 02DEC2024 - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "1) WAS ANY OF THE IMPLANT PROTRUDING OUT OF THE MC WHEN THE DETACHMENT HAPPENED? YES, THE COIL WAS BEING RETRACTED AND WAS HALF IN THE MC AND HALF IN THE VESSEL WHEN THE DETACHMENT HAPPENED. 2) WAS ANY ADDITIONAL DEVICES USED IN ORDER TO MOVE/RETRIEVE THE IMPLANT COIL? YES, THEY DID OPEN A COIL PUSHER TO TRY TO PUSH THE COIL THE REST OF THE WAY IN, HOWEVER IT DIDN'T WORK AND THAT IS WHY THE CHOSE TO FLUSH THE COIL IN. 3) WAS THERE ANY IMPACT TO THE PATIENT FOLLOWING THE IMPLANT COIL BEING FLUSHED INTO THE TREATMENT LOCATION? YES, THE EMBOLIZATION WAS OFF TARGET. LUCKILY, IT WAS NOT IN THE PARENT VESSEL, SO IN THIS CASE A SLIGHTLY OFF TARGET EMBOLIZATION WAS STILL ACCEPTABLE." INCIDENT REPORT FORM SUB

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21027267 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K220699. One submission often covers many models, so check the product and lot details in each record.

  • Z-1042-2024 · Jan 3, 2024 · Open, Classified
    The Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to em

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page