Momentum MIS UU091-11-7540Recalls & reported problems
Ulrich Medical Usa, Inc. · MIS Screw, polyaxial, dia 7.5mm, length 40mm, cannulated, fenestrated · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810076333137), received between Mar 23, 2024 and Mar 23, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture1
- Solder Joint Fracture1
Patient problems
- Failure of Implant1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 23, 2024Fracture; Solder Joint Fracture; Fracture; Solder Joint Fracture
THE REMOVEABLE TAB ON A MIS POLYAXIAL SCREW, BROKE IN THE WRONG LOCATION. ONE SIDE POST REMAINED LONGER THAN INTENDED. THE SURGEON WAS UNAWARE OF THE POTENTIAL PROBLEM. POST OPERATIVE REVISION SURGURY WAS NECESSARY WHEN THE PATIENT COMPLAINED ABOUT PAIN, WHICH WAS DUE TO THE EXTENDED HEIGHT OF THE TAB. IT WAS DISCOVERED THE TAB DID NOT BREAK IN THE REGION IT WAS SUPPOSED TO. THE SURGEON REPORTED THE INSTRUMENT NECESSARY TO BREAK THE TAB CORRECTLY WAS AVAILABLE IN THE SURGICAL SET BUT WAS NOT USED.
Patient problems: Failure of Implant
MAUDE report 18965232 ↗
Other Thoracolumbosacral Pedicle Screw System devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page