Cortera 150100-85050Recalls & reported problems

Xtant Medical Holdings, Inc. · Screw, Poly, Solid, 8.5x50 · Class 2 · Implant

All models of Cortera (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Nov 28, 2023Class IIOpen, Classified

    Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

    Z-0846-2024 · Root cause: Process control · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code NKB (thoracolumbosacral pedicle screw system) under 21 CFR 888.3070 as a Class II device. 21 CFR 888.3070 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Intended to provide immobilization and stabilization of spinal segments in thoracic, lumbar, and sacral spine as an adjunct to fusion.

Common questions about the Cortera 150100-85050

Has the Cortera 150100-85050 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0846-2024 (Class II), began on Nov 28, 2023: Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the Cortera 150100-85050 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Cortera 150100-85050, and is it still sold?

The FDA's device database lists Xtant Medical Holdings, Inc. as the company responsible for it (brand Cortera, model 150100-85050). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page