TriLeap Lower Extremity Anatomic Plating System 22MM Cannulated Cup ReamerRecalls & reported problems
Crossroads Extremity Systems · 22MM Cannulated Cup Reamer · Class 1
All models of TriLeap Lower Extremity Anatomic Plating System (57)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00810112592375), received between Sep 17, 2024 and Mar 12, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Cut4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 12, 2026Failure to Cut
IT WAS REPORTED ON AN UNKNOWN DATE THE REAMERS INSIDE THE TRILEAP SET WERE NOT SHARP AND TOOK A LOT TO REAM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582636 ↗MalfunctionMar 12, 2026Failure to Cut
IT WAS REPORTED THAT THE REAMERS ARE NOT SHARP AND TAKES A LOT TO REAM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582326 ↗MalfunctionSep 17, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED ON SEPTEMBER 3, 2024, WHILE USING THE INSTRUMENTS, SURGEON REPORTED THAT THE REAMERS WERE NOT SHARP. THIS IS A BRAND NEW TRILEAP SET AND HAVE NOT BEEN USED PREVIOUSLY. THE REAMERS WERE USED IN SURGERY BUT DID NOT AFFECT THE OUTCOME OF THE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20238683 ↗MalfunctionSep 17, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT PATIENT UNDERWENT A MTP FUSION CASE USING THE TRI-LEAP SET ON (B)(6) 2024. THERE WAS AN ISSUE WITH THE CUP AND CONE REAMERS; THE REAMERS NOT BEING SHARP ENOUGH. THE SURGEON USED SIX DIFFERENT REAMERS AND SAID ALL WERE DULL AND NOT CUTTING AWAY AT THE BONE LIKE THEY SHOULD BE. THE SURGEON WAS ASKED IF HE FELT LIKE HIS POWER WAS HIGH ENOUGH AND HE ASSURED THAT IT WAS. THE DOCTOR HAD TO THEN MANUALLY REMOVE THE BONE AT THE JOINT WITH A RONGUER. THE SIZE CUP REAMERS USED IN THE CASE WERE 22,20, AND 18. THE SIZE CONE REAMERS USED IN THE CASE WERE 18,16,14. THERE WAS AN ISSUE WITH THE MOST MEDIAL DISTAL PLATE HOLE. MULTIPLE LOCKING SCREWS WOULD NOT LOCK INTO THE PLATE. WHEN PLACING THE PLATE, THREADED OLIVE WIRES AND COMPRESSION FORCEPS WERE USED TO EVENTUALLY GAIN AXIAL COMPRESSION. A COAXIAL DRILL GUIDE SCREWED INTO THE PLATE WAS USED TO DRILL FOR THE DISTAL MEDIAL SCREW. THEN A 2.7 DRILL BIT WAS USED TO DRILL AND MEASURED FOR A 14MM LOCKING SCREW OFF OF THE DRILL BIT. WHEN PUTTING THE SCREW IN BY HAND, THE SURGEON COMPLAINED THAT THE SCREW WAS NOT LOCKING INTO THE PLATE. ANOTHER 14MM LOCKING SCREW WAS TRIED AS WELL AS A 16MM LOCKING SCREW BUT WITH NO LUCK. THE SUR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20236593 ↗
Other Reamer devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page