Skip to content

Ruby XL Coil System RBYPCXL30-ARecalls & reported problems

Penumbra, Inc. · Packing Coil XL, 30cm · Class 2 · Implant

All models of Ruby XL Coil System (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00810169721568), received between Aug 13, 2025 and Jun 2, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Migration” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration3
  • Material Deformation1
  • Physical Resistance/Sticking1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 2, 2026Migration

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE PULMONARY ARTERY USING A PACKING COIL XL, A POD XL AND A NON- PENUMBRA CATHETER (5FR). IT WAS REPORTED THAT THE COILS WOULD NOT STAY AT THE INTENDED LOCATION AND MIGRATED TO ANOTHER BRANCH OF THE PULMONARY ARTERY. THE PHYSICIAN REMOVED BOTH COILS USING A SNARE DEVICE. THE PROCEDURE WAS COMPLETED USING RUBY COILS AND THE SAME CATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25368247 ↗
  • MalfunctionDec 18, 2025Material Deformation; Physical Resistance/Sticking; Premature Separation; Material Deformation; Physical Resistance/Sticking; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE EXTERNAL ILIAC ARTERY USING A PACKING COIL XL AND A NON-PENUMBRA ANGLED DIAGNOSTIC CATHETER. WHILE REPOSITIONING THE PACKING COIL XL, THE PHYSICIAN EXPERIENCED RESISTANCE, AND THE PACKING COIL XL COULD NOT ADVANCE OR RETRACT. WHILE REMOVING THE PACKING COIL XL, IT BECAME DAMAGED BUT REMAINED INTACT TO ITS PUSHER ASSEMBLY. THE PHYSICIAN CONTINUED TO PULL THE PACKING COIL XL WITH FORCE AND KINKED THE PUSHER ASSEMBLY CAUSING THE EMBOLIZATION COIL TO DETACH OUTSIDE OF THE CATHETER. THE EMBOLIZATION COIL COULD NOT BE RETRACTED FROM THE CATHETER AND WAS REMOVED USING A SNARE DEVICE. NO MATERIAL OF THE EMBOLIZATION COIL WAS LEFT IN THE PATIENT¿S BODY. THE PROCEDURE WAS COMPLETED USING A RUBY COIL AND A MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23846313 ↗
  • MalfunctionNov 18, 2025Migration; Migration

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE LEFT MAIN PULMONARY TRUNK USING POD COILS, A POD PACKING COIL (PACKING COIL), PACKING COIL XLS, A NON-PENUMBRA CATHETER, AND A NON-PENUMBRA MICROCATHETER. DURING THE PROCEDURE, THE PHYSICIAN PLACED EIGHT POD COILS, ONE PACKING COIL, AND THREE PACKING COIL XLS IN THE TARGET LOCATION. AFTER PLACING A FINAL PACKING COIL IN THE TARGET LOCATION, THE PHYSICIAN PERFORMED A FLUOROSCOPY AND NOTICED THE COILS PLACED IN THE TARGET SITE HAD SHIFTED. IT WAS REPORTED THE TAIL END OF THE 30CM PACKING COIL XL HAD COME OUT OF THE TARGET LOCATION AND WAS IN AN ADJACENT PULMONARY TRUNK THAT WAS MEDIAL AND LATERAL TO THE TARGET SITE. THE PHYSICIAN USED A SNARE DEVICE TO SUCCESSFULLY REMOVE THE 30CM PACKING COIL XL. THE PROCEDURE HAD ENDED AT THIS POINT. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23587338 ↗
  • MalfunctionAug 13, 2025Migration; Migration

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE SPLENIC ARTERY USING PACKING COIL XL AND A NON-PENUMBRA GLIDE CATHETER (4F). IT WAS REPORTED THAT THE TARGET VESSEL WAS A HIGH FLOW AREA. DURING THE PROCEDURE, THE PHYSICIAN PLACED TWO PACKING COIL XLS IN THE TARGET VESSEL USING THE GLIDE CATHETER. AFTER REMOVING THE GLIDE CATHETER, THE PHYSICIAN NOTICED UNDER FLUOROSCOPY THAT THE PREVIOUSLY PLACED PACKING COIL XL MIGRATED FROM THE SPLENIC ARTERY TO THE HEPATIC ARTERY. A SNARE DEVICE WAS USED TO REMOVE THE PACKING COIL XL FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A RUBY COIL, A POD PACKING COIL (PACKING COIL), AND A MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22794567 ↗