Ruby XL Coil System RBYPCXL70-ARecalls & reported problems

Penumbra, Inc. · Packing Coil XL, 70cm · Class 2 · Implant

All models of Ruby XL Coil System (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00810169721599), received between Oct 27, 2025 and May 13, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation4
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2026Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT GASTRIC VARICES USING A PACKING COIL XL, A NON-PENUMBRA CATHETER. WHILE ADVANCING A PACKING COIL XL THROUGH THE CATHETER, THE PHYSICIAN EXPERIENCED RESISTANCE. WHILE PUSHING THROUGH THE RESISTANCE, THE PACKING COIL XL PUSHER ASSEMBLY KINKED THEN FRACTURED. THIS CAUSED THE EMBOLIZATION COIL TO DETACH PARTIALLY WITHIN THE CATHETER AND THE TARGET VESSEL. THE PHYSICIAN ADVANCED A GUIDEWIRE (.035) TO PUSH THE REMAINING DETACHED EMBOLIZATION COIL INTO THE TARGET LOCATION. AFTER RETRACTING THE CATHETER, THE PHYSICIAN NOTICED UNDER FLUOROSCOPY THAT THE TAIL END OF THE PACKING COIL XL WAS FLOWING INTO THE SPLENIC VEIN. THE PHYSICIAN USED A SNARE DEVICE TO REMOVE THE EMBOLIZATION COIL FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH PACKING COILS AND A NON-PENUMBRA MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25159347 ↗
  • MalfunctionMar 20, 2026Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE LEFT GONADAL VEIN USING RUBY COILS, RUBY COIL XLS, A PACKING COIL XL, AND A NON-PENUMBRA CATHETER. THE PHYSICIAN PLACED RUBY COILS AND RUBY COIL XLS IN THE TARGET LOCATION. THE CATHETER WAS THEN FLUSHED. WHILE ADVANCING A PACKING COIL XL INTO THE TARGET LOCATION, THE PHYSICIAN NOTICED THE EMBOLIZATION COIL WOULD NOT FORM AND PACK IN THE VESSEL. AFTER MAKING SEVERAL ATTEMPTS TO ADVANCE AND FORM THE COIL IN THE VESSEL, THE PHYSICIAN DECIDED TO REMOVE THE PACKING COIL XL. WHILE RETRACTING THE PACKING COIL XL, THE EMBOLIZATION COIL HAD UNINTENTIONALLY DETACHED WITHIN THE CATHETER WITH THE DISTAL END OF THE COIL STICKING OUT INTO THE VESSEL. THE CATHETER CONTAINING THE DETACHED PACKING COIL XL WAS REMOVED AND THE COIL WAS FLUSHED OUT FROM THE CATHETER ON THE BACK TABLE. THE PROCEDURE WAS COMPLETED USING ADDITIONAL RUBY COIL XLS, A POD XL, AND THE SAME CATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24650989 ↗
  • MalfunctionJan 9, 2026Material Deformation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A ESOPHAGEAL VARICES AND SPLENIC RENAL SHUNT USING A PACKING COIL XLS AND A NON-PENUMBRA CATHETER. DURING THE PROCEDURE, THE PHYSICIAN PLACED MULTIPLE PACKING COIL XLS INTO THE TARGET VESSEL. IT WAS NOTED THAT THE PHYSICIAN FLUSHED THE CATHETER BETWEEN EACH COIL. WHILE ADVANCING A NEW PACKING COIL XL, THE COIL WOULD NOT EXIT THE CATHETER. THE PHYSICIAN FLUSHED THE CATHETER AND ATTEMPTED TO RE-ADVANCE THE PACKING COIL XL TWICE WITHOUT SUCCESS. WHILE RE-ADVANCING A THIRD TIME, THE PHYSICIAN KINKED THE PACKING COIL XL PUSHER ASSEMBLY AND THE EMBOLIZATION COIL UNINTENTIONALLY DETACHED WITHIN THE CATHETER. THE CATHETER CONTAINING THE EMBOLIZATION COIL WAS REMOVED. THE PROCEDURE WAS COMPLETED WITH ADDITIONAL PENUMBRA COILS AND THE SAME CATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24028500 ↗
  • MalfunctionOct 27, 2025Premature Separation; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION IN THE OVARIAN VEIN PROCEDURE USING A PACKING COIL XL, A NON-PENUMBRA CATHETER (4F GLIDECATH), AND A GUIDEWIRE. DURING THE PROCEDURE, WHILE ADVANCING A PACKING COIL XL THROUGH THE GLIDE CATHETER, THE EMBOLIZATION COIL UNINTENTIONALLY DETACHED INSIDE THE CATHETER. THE PHYSICIAN USED A GUIDEWIRE TO PUSH THE DETACHED EMBOLIZATION COIL INTO THE TARGET LOCATION. THE PROCEDURE WAS COMPLETED WITH ADDITIONAL COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393809 ↗

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →