Ruby XL Coil System RBYPODXL0655-ARecalls & reported problems

Penumbra, Inc. · POD XL, 6mm x 55cm · Class 2 · Implant

All models of Ruby XL Coil System (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810169721711), received between Nov 13, 2025 and Jan 15, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 15, 2026Unraveled Material

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE VARICES USING A POD XL, A NON-PENUMBRA COIL, AND A NON-PENUMBRA CATHETER. DURING THE PROCEDURE, THE PHYSICIAN SUCCESSFULLY PLACED ONE NON-PENUMBRA COIL IN THE TARGET LOCATION. WHILE ADVANCING A POD XL, THE PHYSICIAN NOTICED THAT THE POD XL WAS TOO LARGE FOR THE TARGET VESSEL. WHILE RETRACTING THE POD XL, THE PHYSICIAN KINKED THE POD XL PUSHER ASSEMBLY. THE PHYSICIAN CONTINUED TO RETRACT THE POD XL AND NOTICED THAT THE POD XL UNRAVELED. THE POD XL WAS ABLE TO BE COMPLETELY REMOVED. THE PROCEDURE WAS COMPLETED AT THIS POINT. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24077400 ↗
  • MalfunctionNov 13, 2025Migration; Migration

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE RIGHT INTERNAL ILIAC ARTERY (IIA) USING POD XLS, A NON-PENUMBRA BRAIDED CATHETER, AND NON-PENUMBRA SHEATHS. DURING THE PROCEDURE, THE PHYSICIAN SUCCESSFULLY PLACED A POD XL INTO THE TARGET LOCATION. WHILE RETRACTING THE CATHETER, THE PHYSICIAN OBSERVED UNDER FLUOROSCOPY THE POD XL MOVED INTO AN ADJACENT ACCESSORY ARTERY. A SNARE DEVICE WAS USED TO REPOSITION THE POD XL BUT WAS UNSUCCESSFUL. WHILE RETRACTING THE SNARE DEVICE, THE POD XL BECAME DAMAGED. THE PHYSICIAN THEN DECIDED TO REMOVE THE EMBOLIZATION COIL USING THE SNARE. NO PART OF THE POD XL WAS LEFT IN THE PATIENT. THE PHYSICIAN THEN ADVANCED A NEW POD XL TO THE TARGET LOCATION AND ATTEMPTED TO DETACH THE COIL. HOWEVER, THE EMBOLIZATION COIL WAS STILL ATTACHED TO THE PUSHER ASSEMBLY. THE PHYSICIAN EXPERIENCED RESISTANCE REMOVING THE POD XL. THE POD XL WAS SUCCESSFULLY REMOVED. THE EMBOLIZATION COIL WAS DAMAGED UPON REMOVAL. THE PROCEDURE WAS COMPLETED USING A NON-PENUMBRA DEVICE, THE SAME BRAIDED CATHETER, AND THE SAME SHEATHS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23544193 ↗

Other Device, Vascular, For Promoting Embolization devices