Skip to content

Ruby XL Coil System RBYPODXL1055-ARecalls & reported problems

Penumbra, Inc. · POD XL, 10mm x 55cm · Class 2 · Implant

All models of Ruby XL Coil System (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810169721742), received between Nov 10, 2025 and Jun 2, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation2
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 2, 2026Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN TWO VARICES OF THE LEFT RENAL VEIN USING A POD XL COIL (POD XL), A NON-PENUMBRA CATHETER (4FR), AND A NON-PENUMBRA SHEATH (7FR). THE PHYSICIAN SUCCESSFULLY PLACED MULTIPLE COILS IN THE FIRST VARIX. THE CATHETER WAS THEN REPOSITIONED TO THE SECOND VARIX AND PLACED ONE NON-PENUMBRA COIL. WHILE ADVANCING THE POD XL THROUGH THE CATHETER, THE PHYSICIAN EXPERIENCED RESISTANCE CAUSING THE COIL'S PUSHER ASSEMBLY TO KINK. SUBSEQUENTLY, THE POD XL UNINTENTIONALLY DETACHED WITHIN THE CATHETER. THE CATHETER WAS REMOVED; HOWEVER, THE EMBOLIZATION COIL REMAINED WITHIN THE SHEATH. THE SHEATH CONTAINING THE EMBOLIZATION COIL WAS REMOVED FROM THE PATIENT. ACCESS WAS LOST AND THE PROCEDURE WAS CONSIDERED COMPLETED. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25367141 ↗
  • MalfunctionNov 10, 2025Physical Resistance/Sticking; Premature Separation; Physical Resistance/Sticking; Premature Separation

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE VARICES USING A POD XL AND A NON-PENUMBRA CATHETER. IT WAS NOTED THAT THE PATIENT¿S ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, THE PHYSICIAN OBTAINED DIRECT STICK ACCESS IN THE ABDOMEN. THE POD XL WAS ADVANCED TO THE DISTAL END OF THE CATHETER, BUT WHILE ATTEMPTING TO ADVANCE THE POD XL IN THE VEIN, THE PHYSICIAN WAS EXPERIENCING RESISTANCE AND DECIDED TO REMOVE THE POD XL. THEN, WHILE REMOVING THE POD XL, THE COIL UNINTENTIONALLY DETACHED INSIDE THE CATHETER. THEREFORE, THE CATHETER CONTAINING THE DETACHED POD XL WAS REMOVED TOGETHER AS A UNIT. THE PROCEDURE WAS COMPLETED USING NEW POD XLS, PACKING COIL XLS, POD COILS, AND A NEW CATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23513535 ↗

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates