Azur Detachable Framing Coil 45-651434Recalls & reported problems

Microvention, Inc. · Azur · Class 2 · Implant

All models of Azur Detachable Framing Coil (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00810170016127), received between Oct 27, 2017 and Jul 9, 2024 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component4
  • Detachment Of Device Component1
  • Device-Device Incompatibility1
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 9, 2024Separation Problem; Separation Problem

    AS REPORTED: DURING A PROXIMAL SPLENIC EMBO, DR. HAD A 30CM SHEATH IN THE AORTA AND A 100CM 4FR ANGLED CATHETER IN THE SPLENIC ARTERY. THE ARTERY MEASURED 8-9MM IN DIAMETER. THE FIRST COIL USED WAS AN 35 14X34 FRAMING COIL. DR. ADVANCED THE COIL TO THE TIP OF THE CATHETER AND IT BECAME STUCK. WITHIN A FEW SECONDS OF MANIPULATING THE COIL, DR. STATED THAT HE FELT LIKE THE COIL BROKE FREE. HE REMOVED WHAT WAS LEFT OF THE PUSHER WIRE AND INJECTED SALINE INTO THE CATHETER TO SHOOT THE COIL OUT INTO THE VESSEL. THE COIL LANDED WHERE HE WANTED, HOWEVER, A LONG PORTION OF THE PUSHER WIRE WAS STILL ATTACHED TO THE COIL. THE WIRE SNAPPED IN HALF, IT WASN'T THE COIL DETACHING LIKE HE SUSPECTED. THE WIRE PROLAPSED INTO A COLLATERAL VESSEL AND TRACKED ALL THE WAY BACK TO THE ORIGIN OF THE SPLENIC ARTERY. IT WAS NOTICED THAT THE CATHETER KINKED AT THE PROXIMAL PORTION OF THE SPLENIC ARTERY. FROM THERE, HE DECIDED TO CONTINUE THE PROCEDURE AND PUT A 45CM SHEATH UP INTO THE SPLENIC ARTERY ALONG WITH A NEW 4FR 100CM ANGLED CATHETER AND CONTINUED TO DEPLOY SIX 035 AZUR CX COILS WITHOUT INCIDENT TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19707223 ↗
  • MalfunctionMar 15, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    PLEASE SEE H10 FOR DEVICE INVESTIGATION RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16550757 ↗
  • MalfunctionJun 24, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING TREATMENT OF A PELVIC CONGESTION CASE, A COMBINATION OF MULTIPLE EMBOLIZATION COIL IMPLANTS WAS USED WHEN THE FRAMING COIL UNEXPECTEDLY DETACHED IN THE MICROCATHETER. THE COIL, TOGETHER WITH THE CATHETER, WERE REMOVED AND REPLACED TO COMPLETE THE CASE. THERE WAS NO REPORTED PATIENT INJURY OR ADDITIONAL INTERVENTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12063042 ↗
  • MalfunctionJun 4, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING TREATMENT OF AN ANEURYSM, AN EMBOLIZATION COIL IMPLANT UNEXPECTEDLY DETACHED IN THE MICROCATHETER. THE COIL, TOGETHER WITH THE CATHETER, WERE REMOVED FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY OR ADDITIONAL INTERVENTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11942552 ↗
  • MalfunctionDec 20, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT AN EMBOLIZATION IMPLANT COIL UNEXPECTEDLY DETACHED IN A MICROCATHETER DURING TREATMENT OF AN ENDOLEAK IN THE INTERNAL ILIAC ARTERY. THE COIL WAS REMOVED IN ITS ENTIRETY TOGETHER WITH MICROCATHETER. THERE WAS NO REPORTED PATIENT INJURY OR INTERVENTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9508080 ↗
  • MalfunctionOct 27, 2017Detachment Of Device Component; Device-Device Incompatibility; Detachment Of Device Component; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING THE TREATMENT OF A DISTAL SPLENIC ANEURYSM, THE COIL GOT STUCK AT THE TIP OF THE MICROCATHETER. WHEN THE COIL WAS PUSHED HARDER, THE PUSHER WIRE SEPARATED FROM THE COIL AND THE COIL THEN DETACHED IN THE MICROCATHETER. AN ATTEMPT WAS MADE TO PUSH THE COIL INTO THE ANEURYSM WITH A GUIDEWIRE; HOWEVER, THE COIL REMAINED IN THE MICROCATHETER. BOTH SEGMENTS OF THE COIL WERE WITHDRAWN TOGETHER WITH THE MICROCATHETER, AND REMOVED IN THEIR ENTIRETY. THERE WAS NO REPORTED INTERVENTION OR PATIENT INJURY. THE PATIENT IS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6985631 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →