1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00810170018862), received between Apr 7, 2023 and Nov 6, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2024Class IIOpen, Classified

    Due to Retrieval Device product labels stating the incorrect expiration dates that (5 years instead of 3 years) which extends beyond their true expiration date.

    Z-3157-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Separation2
  • Detachment of Device or Device Component1
  • Retraction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    NO NEW INFORMATION WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23490940 ↗
  • InjuryOct 17, 2024Material Separation; Material Separation

    AS REPORTED: AN ERIC HAD BROKEN AND WAS IN THE PATIENT WHILE A PHYSICIAN WAS ATTEMPTING A STROKE THROMBECTOMY. FORTY-FIVE MINUTES LATER, THEY WERE STILL ATTEMPTING TO RETRIEVE THE ERIC WITH A NUMBER OF DEVICES: STENTRIEVERS, SNARES, ETC. IT TOOK THEM OVER AN HOUR, BUT EVENTUALLY, THE PHYSICIAN WAS ABLE TO USE A NON-MVI CLOT RETRIEVER TO EXTRACT THE ERIC. THE PHYSICIAN USED NUMEROUS RETRIEVAL ATTEMPTS TO EVENTUALLY GET THE ERIC OUT OF THE VESSEL. UPON PHYSICAL INSPECTION, IT APPEARS THE ERIC SHEARED OFF AT THE JOINT WHERE THE PUSHER WIRE IS CONNECTED TO THE ERIC CAGES. VESSEL WAS TORTUOUS. THE NON-MVI CLOT RETRIEVER THAT WAS USED TO CAPTURE THE ERIC IS STILL ENTWINED. THE PHYSICIAN SAID IT WAS A DISTAL M1 BORDERING M2. UNSURE ON CLOT TYPE AS THEY WERE UNABLE TO EXTRACT THE OCCLUSION. THERE WAS ONE PASS PRIOR TO EVENT. PATIENT WAS SENT TO NEURO ICU. THEY WERE UNABLE TO REMOVE OCCLUSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20476931 ↗
  • MalfunctionApr 7, 2023Retraction Problem; Material Separation; Retraction Problem; Material Separation

    IT WAS REPORTED THAT DURING TREATING A RIGHT M1 STROKE WITH A CLOT RETRIEVER. A MICROCATHETER AND A GUIDE CATHETER WERE USED. AFTER THE FIRST PASS OF THE ERIC THE TECH CLEANED IT AND TRIED PULLING IT BACK THROUGH GUIDE CATHETER, EXPERIENCED FRICTION, AND THE DEVICE COULD NOT BE RETRACTED. THE ENTIRE SYSTEM WAS REMOVED AND THE CASE WAS COMPLETED. THE PHYSICIAN TRIED TO PULL THE CLOT RETRIEVER THROUGH THE GUIDE CATHETER ON THE BACK TABLE AND THE WIRE SEPARATED FROM THE DEVICE. THERE WAS NO REPORTED HARM OR INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16700707 ↗

FDA clearance and classification

Cleared through 510(k) K211120 on Mar 31, 2022 (decision: Substantially Equivalent). Submitted by MicroVention, Inc. as “ERIC Retrieval Device”. The FDA review took 350 days. FDA record ↗

The FDA regulates product code NRY (catheter, thrombus retriever) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗

FDA definition of this device type: The device is intended to restore blood flow by removing thrombus/clots in patients experiencing ischemic stroke. The prior clearances in this category have been limited to simply identifying catheter placed in the peripheral, coronary, and neurovasculature. It is felt that this is a unique claim and should not be combined with previously cleared catheters under a general procode.

Common questions about the ERIC Retrieval Device ER176044-US

Has the ERIC Retrieval Device ER176044-US been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3157-2024 (Class II), began on Aug 5, 2024: Due to Retrieval Device product labels stating the incorrect expiration dates that (5 years instead of 3 years) which extends beyond their true expiration date. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ERIC Retrieval Device ER176044-US?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Apr 7, 2023 and Nov 6, 2025. In those reports, the most common specific device problems are “Material Separation” (2), “Detachment of Device or Device Component” (1) and “Retraction Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ERIC Retrieval Device ER176044-US?

None of the reports are classified as deaths, 2 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ERIC Retrieval Device ER176044-US FDA approved?

It was cleared through 510(k) K211120 on Mar 31, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the ERIC Retrieval Device ER176044-US, and is it still sold?

The FDA's device database lists Microvention, Inc. as the company responsible for it (brand ERIC Retrieval Device, model ER176044-US). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Thrombus Retriever devices

All Catheter, Thrombus Retriever devices →