ERIC Retrieval DeviceRecalls & reported problems across all models
Microvention, Inc. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one ERIC Retrieval Device model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 1
We found 3 models of ERIC Retrieval Device by Microvention, Inc. in the FDA's device registry (GUDID). Together they are named in 5 adverse event reports in the FDA's MAUDE database, received between Apr 7, 2023 and Nov 6, 2025 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (3 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
ERIC Retrieval Device recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Aug 5, 2024Open, Classified
Due to Retrieval Device product labels stating the incorrect expiration dates that (5 years instead of 3 years) which extends beyond their true expiration date.
Z-3157-2024 · Root cause: Under Investigation by firm · names 3 of these models · Recall details · FDA record ↗
Most reported ERIC Retrieval Device problems
Device problems
- Material Separation3
- Break1
- Detachment of Device or Device Component1
- Retraction Problem1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Foreign Body In Patient1
- Stroke/CVA1
ERIC Retrieval Device models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- ERIC Retrieval Device ER176044-USCatheter, Thrombus Retriever1 recall3 reports
- ERIC Retrieval Device ER173020-USCatheter, Thrombus Retriever1 recall1 report
- ERIC Retrieval Device ER174030-USCatheter, Thrombus Retriever1 recall1 report
Recent ERIC Retrieval Device adverse event reports
The 5 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryNov 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
NO NEW INFORMATION WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23490940 ↗InjuryDec 30, 2024Break; Break
ADDITIONAL INFORMATION WAS PROVIDED THAT STATED THE PATIENT'S ANATOMY WAS TORTUOUS WHICH MADE IT DIFFICULT TO REMOVE THE DEVICE, AND IT BROKE. THE PATIENT'S CONDITION REMAINS THE SAME AS STROKE SCALE OF NIHSS 20. NO FURTHER INFORMATION IS AVAILABLE FOR THIS CASE.
Patient problems: Foreign Body In Patient; Stroke/CVA
MAUDE report 21039650 ↗InjuryOct 17, 2024Material Separation; Material Separation
AS REPORTED: AN ERIC HAD BROKEN AND WAS IN THE PATIENT WHILE A PHYSICIAN WAS ATTEMPTING A STROKE THROMBECTOMY. FORTY-FIVE MINUTES LATER, THEY WERE STILL ATTEMPTING TO RETRIEVE THE ERIC WITH A NUMBER OF DEVICES: STENTRIEVERS, SNARES, ETC. IT TOOK THEM OVER AN HOUR, BUT EVENTUALLY, THE PHYSICIAN WAS ABLE TO USE A NON-MVI CLOT RETRIEVER TO EXTRACT THE ERIC. THE PHYSICIAN USED NUMEROUS RETRIEVAL ATTEMPTS TO EVENTUALLY GET THE ERIC OUT OF THE VESSEL. UPON PHYSICAL INSPECTION, IT APPEARS THE ERIC SHEARED OFF AT THE JOINT WHERE THE PUSHER WIRE IS CONNECTED TO THE ERIC CAGES. VESSEL WAS TORTUOUS. THE NON-MVI CLOT RETRIEVER THAT WAS USED TO CAPTURE THE ERIC IS STILL ENTWINED. THE PHYSICIAN SAID IT WAS A DISTAL M1 BORDERING M2. UNSURE ON CLOT TYPE AS THEY WERE UNABLE TO EXTRACT THE OCCLUSION. THERE WAS ONE PASS PRIOR TO EVENT. PATIENT WAS SENT TO NEURO ICU. THEY WERE UNABLE TO REMOVE OCCLUSION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20476931 ↗MalfunctionMay 3, 2024Material Separation; Stretched; Material Separation; Stretched
IT WAS REPORTED THAT THE CLOT RETRIEVER DEVICE WAS USED DURING A STROKE CASE. REPORTEDLY, AFTER USE AND REMOVAL FROM THE PATIENT AND ON THE BACK TABLE, IT WAS NOTED THAT THE DISTAL 5MM TIP OF THE LEAD WIRE STRETCHED AND CAME APART AT SOME POINT AFTER SECOND PASS. THE CASE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE OF THE SAME MODEL. THE PATIENT IS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19238204 ↗MalfunctionApr 7, 2023Retraction Problem; Material Separation; Retraction Problem; Material Separation
IT WAS REPORTED THAT DURING TREATING A RIGHT M1 STROKE WITH A CLOT RETRIEVER. A MICROCATHETER AND A GUIDE CATHETER WERE USED. AFTER THE FIRST PASS OF THE ERIC THE TECH CLEANED IT AND TRIED PULLING IT BACK THROUGH GUIDE CATHETER, EXPERIENCED FRICTION, AND THE DEVICE COULD NOT BE RETRACTED. THE ENTIRE SYSTEM WAS REMOVED AND THE CASE WAS COMPLETED. THE PHYSICIAN TRIED TO PULL THE CLOT RETRIEVER THROUGH THE GUIDE CATHETER ON THE BACK TABLE AND THE WIRE SEPARATED FROM THE DEVICE. THERE WAS NO REPORTED HARM OR INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16700707 ↗
Other Microvention, Inc. products
- Azur CX Detachable22 models · 509 reports · 1 recall
- Stent63 models · 237 reports · 2 recalls
- Azur Detachable Helical HydroCoil42 models · 190 reports
- V-Trak75 models · 147 reports
- WEB14 models · 134 reports
- AZUR14 models · 104 reports
- HydroSoft 3D26 models · 101 reports
- MicroPlex Helical Soft9 models · 8 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page