Prometra Patient Therapy Controller 12860Recalls & reported problems
Flowonix Medical, Inc. · The Patient Therapy Controller (PTC) is a handheld touchscreen device, which allows a patient to initiate a pre-configured supplemental bolus of medication from their implanted Programmable Pump. The patient requests a bolus by pressing the Rx button on the PTC and then placing it over their implanted pump. The Patient Therapy Controller communicates with the implanted pump initiating the supplemental bolus delivery. The settings for the bolus delivery are programmed into the Patient Therapy Controller by a healthcare certified professional (HCP) using the Configuration Device. · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
The Flowonix Prometra Patient Therapy Controller is a handheld touchscreen device that patients hold over their implanted pump to request a pre-set extra dose of medication. The 58 reports mostly describe the controller not working as expected — error codes, wrong messages on the screen, batteries draining fast or erased prescription settings, buttons not responding, and one cracked screen after being dropped. Three reports describe injuries, including symptoms of too much medication delivery; no deaths were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 58
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 58 adverse event reports that list this exact device (DI 00810335020174), received between Nov 5, 2015 and Dec 6, 2019 — classified as 3 injuries, 55 malfunctions.
The most frequently reported specific device problem is “Device Operational Issue” (32 reports).
13 other recalls cover devices cleared under the same FDA submission (P080012); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Operational Issue32
- Device Displays Incorrect Message15
- Device Issue5
- Device Operates Differently Than Expected3
- Mechanical Problem3
- Communication or Transmission Problem2
- Crack2
- Material Integrity Problem2
Patient problems
- No Known Impact Or Consequence To Patient23
- No Consequences Or Impact To Patient16
- No Patient Involvement12
- Patient Problem/Medical Problem3
- Pain2
- Inadequate Pain Relief1
- Insufficient Information1
- Nausea1
Recent adverse event reports
The 25 most recent of 58 reports, in the reporter's own words as published by the FDA.
InjuryDec 6, 2019Excess Flow or Over-Infusion; Adverse Event Without Identified Device or Use Problem; Excess Flow or Over-Infusion; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD MULTIPLE ERROR CODE 38S AND POTENTIAL OVER DELIVERY OF MEDICATION. PATIENT REPORTS WITHDRAWAL SYMPTOMS - SWEATING, NAUSEA AND VOMITING.
Patient problems: Nausea; Patient Problem/Medical Problem
MAUDE report 9433081 ↗InjuryOct 13, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
A HEALTH CARE PROFESSIONAL REPORTED THAT THE PATIENT WAS NOT ADMITTED TO THE HOSPITAL. THE PATIENT HAD ONLY VISITED THE EMERGENCY ROOM AND WAS LATER RELEASED. NO ADDITIONAL ISSUES WERE REPORTED.
Patient problems: Patient Problem/Medical Problem; Vomiting
MAUDE report 6949858 ↗MalfunctionDec 15, 2016Device Operational Issue; Device Operational Issue
IT WAS REPORTED THAT A BUTTON ON THE PATIENT THERAPY CONTROLLER (PTC) IS NOT RESPONSIVE UNLESS THE DEVICE IS PLUGGED INTO A WALL CHARGER AND UNPLUGGED. THERE WERE NO PATIENT EFFECTS REPORTED, AND THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6179433 ↗MalfunctionDec 15, 2016Device Operational Issue; Device Operational Issue
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) WOULD NOT TURN ON. THERE WERE NO PATIENT EFFECTS REPORTED, AND THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6179315 ↗MalfunctionDec 15, 2016Device Issue; Device Issue
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) BATTERY DEPLETED QUICKLY, AND THAT THE PRESCRIPTION INFORMATION ON THE PTC WAS ERASED TWICE. THE PATIENT REPORTED EXPERIENCING INCREASED PAIN. THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: Pain
MAUDE report 6179242 ↗MalfunctionDec 9, 2016Device Issue; Device Issue
IT WAS REPORTED THE PATIENT THERAPY CONTROLLER (PTC) WAS INCONSISTENT IN BOLUS DELIVERY, AND DISPLAYED BATTERY CHARGING ISSUES. THERE WERE NO PATIENT EFFECTS REPORTED, AND THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6157812 ↗MalfunctionDec 2, 2016Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) WAS RECONFIGURED DURING AN APPOINTMENT. AFTER THE RECONFIGURATION WAS COMPLETE, THE PTC DISPLAYED A MESSAGE INDICATING THAT THE DEVICE MUST BE CONFIGURED BEFORE USE MULTIPLE TIMES OVER A THREE WEEK PERIOD. THERE WERE NO PATIENT EFFECTS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6142614 ↗MalfunctionDec 2, 2016Crack; Device Operational Issue; Crack; Device Operational Issue
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) WAS DROPPED, WHICH CAUSED THE SCREEN TO CRACK. THE DEVICE SUBSEQUENTLY COULD NOT BE POWERED ON. THERE WERE NO PATIENT EFFECTS REPORTED, AND THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6142611 ↗MalfunctionDec 2, 2016Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) DISPLAYED BATTERY CHARGING AND BOLUS DELIVERY ISSUES. THERE WERE NO PATIENT EFFECTS REPORTED, AND THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6142603 ↗MalfunctionNov 23, 2016Device Issue; Device Issue
IT WAS REPORTED THE PATIENT THERAPY CONTROLLER (PTC) WAS INCONSISTENT IN BOLUS DELIVERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6125414 ↗MalfunctionNov 23, 2016Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) COULD NOT BE POWERED ON, BUT DISPLAYED THAT IT WAS FULLY CHARGED WHEN CONNECTED TO THE CHARGER. THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6125399 ↗MalfunctionNov 9, 2016Device Operational Issue; Device Operational Issue
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) BECAME UNRESPONSIVE. TROUBLESHOOTING ATTEMPTS WERE NOT SUCCESSFUL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6091060 ↗MalfunctionNov 9, 2016Crack; Device Operational Issue; Crack; Device Operational Issue
IT WAS REPORTED THAT THE TOUCHSCREEN ON THE PATIENT THERAPY CONTROLLER (PTC) WAS NOT FUNCTIONAL, AND THAT THE SCREEN HAS A HAIRLINE CRACK.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6091059 ↗MalfunctionNov 2, 2016Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) POWER BUTTON IS WORN AND DIFFICULT TO USE. THERE WERE NO PATIENT EFFECTS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6074502 ↗MalfunctionNov 2, 2016Device Operational Issue; Device Operational Issue
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) BECAME UNRESPONSIVE. TROUBLESHOOTING ATTEMPTS WERE NOT SUCCESSFUL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 6074486 ↗MalfunctionOct 19, 2016Device Operational Issue; Device Operational Issue
IT WAS REPORTED THAT THE BUTTONS ON THE PATIENT THERAPY CONTROLLER (PTC) WERE DAMAGED. THE DEVICE WAS UNRESPONSIVE WHEN CHARGED OR WHEN A BUTTON WAS PUSHED. THE PATIENT REPORTED THAT THERE WAS NO PHYSICAL DAMAGE. THE PATIENT REPORTED THAT SHE EXPERIENCED A DECREASE IN PAIN RELIEF. THE MANAGING INFUSION NURSE LATER REPORTED THAT THE BUTTON ISSUE DID NOT APPEAR TO BE DUE TO WEAR AND TEAR DUE TO NORMAL USAGE, AND THAT IT IS POSSIBLE THE DEVICE WAS DAMAGED IN SOME WAY.
Patient problems: Pain
MAUDE report 6039656 ↗MalfunctionOct 19, 2016Device Operational Issue; Device Operational Issue
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) BECAME UNRESPONSIVE. TROUBLESHOOTING ATTEMPTS WERE NOT SUCCESSFUL. THERE WERE NO PATIENT EFFECTS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6039655 ↗MalfunctionOct 12, 2016Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A BUTTON ON THE PATIENT THERAPY CONTROLLER (PTC) WAS DAMAGED FROM USE. THERE WERE NO PATIENT EFFECTS REPORTED. THE DEVICE IS FUNCTIONAL AND WILL NOT BE RETURNED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6023949 ↗MalfunctionOct 6, 2016Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER WAS RECONFIGURED DURING AN APPOINTMENT. AFTER THE RECONFIGURATION WAS COMPLETE, THE PTC WAS REMOVED FROM THE CHARGER. THE PTC THEN DISPLAYED A MESSAGE INDICATING THAT THE DEVICE MUST BE CONFIGURED BEFORE USE. TROUBLESHOOTING WAS PERFORMED AND THE ISSUE WAS RESOLVED. THERE WERE NO PATIENT EFFECTS REPORTED, AND THE DEVICE WILL NOT BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 6003165 ↗MalfunctionSep 30, 2016Display or Visual Feedback Problem; Device Displays Incorrect Message; Device Operational Issue; Display or Visual Feedback Problem; Device Displays Incorrect Message; Device Operational Issue
AFTER A REFILL APPOINTMENT ON (B)(6) 2016, IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) DISPLAYED AN ERROR MESSAGE INDICATING THAT A PHYSICIAN MUST CONFIGURE THE DEVICE BEFORE USE. THE DEVICE WAS RECONFIGURED ON (B)(6) 2016. DURING THE RECONFIGURATION APPOINTMENT, AN ERROR CODE WAS DISPLAYED WHEN THE PATIENT INITIALLY ATTEMPTED TO DELIVER A BOLUS WITH THE PTC. THE SECOND ATTEMPT WAS SUCCESSFUL. WHEN THIS BOLUS WAS BEING DELIVERED, THE PTC INCORRECTLY DISPLAYED THAT THE NEXT BOLUS WILL BE AVAILABLE IN 20 MINUTES, WHEN IT SHOULD HAVE BEEN 4 HOURS AND 20 MINUTES. THE PATIENT BOLUSED THEMSELVES AGAIN ONCE THE 20 MINUTE PERIOD WAS COMPLETE. THE PTC THEN DISPLAYED THE CORRECT PERIOD OF 4 HOURS AND 20 MINUTES FOR THE NEXT BOLUS. IT WAS LATER REPORTED THAT THE PATIENT WAS ABLE TO GIVE THEMSELVES MORE BOLUS THAN ALLOTTED WITHIN A 24 HOUR PERIOD. THERE WERE NO PATIENT EFFECTS REPORTED. NO FURTHER ISSUES WERE REPORTED, AND THE DEVICE WILL NOT BE RETURNED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5990253 ↗MalfunctionSep 29, 2016Device Issue; Device Issue
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) MAY HAVE BEEN DAMAGED DURING A TORNADO, AND THAT THE REFILL DATES FREQUENTLY CHANGED AFTER THE TORNADO. THE PATIENT NOTED THAT SHE WAS NOT RECEIVING ADEQUATE PAIN RELIEF. THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: Inadequate Pain Relief
MAUDE report 5988835 ↗MalfunctionSep 29, 2016Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A BUTTON ON THE PATIENT THERAPY CONTROLLER (PTC) WAS DAMAGED FROM USE. THERE WERE NO PATIENT EFFECTS REPORTED, AND THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5988825 ↗MalfunctionSep 21, 2016Device Operational Issue; Device Operational Issue
IT WAS REPORTED THAT DURING INITIAL SETUP, THE PATIENT THERAPY CONTROLLER (PTC) WAS ABLE TO SCAN AND CONNECT TO A PROGRAMMER WHILE CHARGING. UPON UNPLUGGING THE PTC FROM THE CHARGER, THE PTC SCREEN WENT BLACK AND BECAME UNRESPONSIVE. TROUBLESHOOTING ATTEMPTS WERE NOT SUCCESSFUL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 5967613 ↗MalfunctionSep 21, 2016Device Operational Issue; Device Operational Issue
IT WAS REPORTED THAT DURING INITIAL SETUP, THE PATIENT THERAPY CONTROLLER (PTC) WAS ABLE TO SCAN AND CONNECT TO A PROGRAMMER WHILE CHARGING. UPON UNPLUGGING THE PTC FROM THE CHARGER, THE PTC SCREEN WENT BLACK AND BECAME UNRESPONSIVE. TROUBLESHOOTING ATTEMPTS WERE NOT SUCCESSFUL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 5967591 ↗MalfunctionSep 21, 2016Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE PATIENT THERAPY CONTROLLER (PTC) WAS RECONFIGURED DURING AN APPOINTMENT. AFTER THE RECONFIGURATION WAS COMPLETE, THE PTC WAS REMOVED FROM THE CHARGER. THE PTC THEN DISPLAYED A MESSAGE INDICATING THAT THE DEVICE MUST BE CONFIGURED BEFORE USE. TROUBLESHOOTING WAS PERFORMED AND THE ISSUE WAS RESOLVED. THERE WERE NO PATIENT EFFECTS REPORTED, AND THE DEVICE WILL NOT BE RETURNED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5967585 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P080012. One submission often covers many models, so check the product and lot details in each record.
- Z-1996-2021 · May 26, 2021 · TerminatedPhysician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Mode
- Z-1124-2020 · Dec 24, 2019 · Open, ClassifiedPrometra Programmable 20mL Pump. Catalog No. 91827 The device is an implantable infusion pump. GTIN-14 00810335020XXX (XXX = in-country language specific)
- Z-1125-2020 · Dec 24, 2019 · Open, ClassifiedPrometra II Programmable 20mL Pump. Catalog No. 93827 The device is an implantable infusion pump. GTIN-14 00810335020YYY (YYY= in-country language specific)
- Z-1123-2020 · Dec 24, 2019 · Open, ClassifiedUS Clinical 20mL Programmable Pump. Catalog No. 01827 The device is an implantable infusion pump. GTIN-14 00810335020082
- Z-1120-2020 · Dec 24, 2019 · Open, ClassifiedPrometra Programmable 20mL Pump. Catalog No. 11827 The device is an implantable infusion pump. GTIN-14 00810335020082
- Z-1122-2020 · Dec 24, 2019 · Open, ClassifiedPrometra II 40mL Pump. Catalog No. 16827 The device is an implantable infusion pump. GTIN-14 00810335020242
- Z-1121-2020 · Dec 24, 2019 · Open, ClassifiedPrometra II 20mL Pump. Catalog No. 13827 The device is an implantable infusion pump. GTIN-14 00810335020228
- Z-1714-2019 · Mar 21, 2019 · TerminatedPrometra Programmable Pump, Cat. No. 11827 For intrathecal infusion.
- Z-1712-2019 · Mar 21, 2019 · TerminatedPrometra Clinician Programmer (Touchscreen) Software version 2.00.29, Cat. No. 12828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion.
- Z-1715-2019 · Mar 21, 2019 · TerminatedPrometra II Programmable Pump, Cat. No. 13827 For intrathecal infusion.
- Z-1716-2019 · Mar 21, 2019 · TerminatedPrometra II Programmable Pump (Clinical), Cat. No. 01827 For intrathecal infusion.
- Z-1713-2019 · Mar 21, 2019 · TerminatedPrometra Programmer (Grand Prime) Software version 2.00.29, Cat. No. 13828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion.
- Z-2740-2017 · May 22, 2017 · TerminatedPROMETRA II Programmable Infusion Pump, Implantable Infusion Pump, Catalog Number/REF 13827, Part Number PL-31790-02. Manufactured by: Flowonix Medical Inc., --- The correction i
Other Pump, Infusion, Implanted, Programmable devices
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall7 deaths4,929 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls5 deaths4,303 reports
- SYNCHROMED ® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths4,119 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls10 deaths3,054 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths3,084 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls1 death2,588 reports
- SynchroMed® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths2,291 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall3 deaths2,242 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page