Lantern Surgical Assistant (HSA) 405204Recalls & reported problems
Orthalign, Inc. · Reference Sensor (RS), LSA · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00810832031451), received between Jul 9, 2024 and Nov 18, 2025 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K200892); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Communication or Transmission Problem1
- Incorrect Measurement1
- Off-Label Use1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Joint Laxity1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 18, 2025Off-Label Use; Communication or Transmission Problem; Off-Label Use; Communication or Transmission Problem
THE REFERENCE SENSOR WOULD NOT CONNECT TO THE LANTERN UNIT. THEY TRIED OPENING A NEW UNIT AND USING IT WITH A NEW BATTERY / OLD BATTERY / ETC. TWO TO THREE YELLOW FLASHES FOLLOWED BY A STEADY YELLOW LIGHT. A DEMO REFERENCE SENSOR WAS USED ON THE PATIENT AND IT FAILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23585490 ↗InjuryAug 26, 2025Incorrect Measurement; Incorrect Measurement
THE LANTERN UNIT'S ACCURACY WAS CALLED INTO QUESTION FOR THE FEMORAL CUT. THE SURGEON CORRECTED INTRA-OPERATIVELY.
Patient problems: Joint Laxity
MAUDE report 22895835 ↗MalfunctionJul 9, 2024Output Problem; Output Problem
TIBIA MEASUREMENTS LOOKED GREAT, BUT WAS UNABLE TO GET FLEXION NUMBERS FOR THE FEMUR. CASE WAS COMPLETED USING THE ORTHALIGN REFERENCE SENSOR FOR THE TIBIA AND STANDARD INSTRUMENTATION FOR THE FEMUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19702853 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K200892. One submission often covers many models, so check the product and lot details in each record.
- Z-1630-2026 · Feb 10, 2026 · Open, ClassifiedDrill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System
- Z-1629-2026 · Feb 10, 2026 · Open, ClassifiedDrill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System
Other Orthopedic Stereotaxic Instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page