Skip to content

Lantern Surgical Assistant (LSA) KA406000-02Recalls & reported problems

Orthalign, Inc. · Packaged Assembly, Nav Unit, LSA (JP) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810832031963), received between Dec 26, 2024 and Dec 26, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K200892); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 26, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE VALUES SHOWN BY THE NAVIGATION UNIT DO NOT MATCH THE ACTUAL VALUES MEASURED. THE EXTENSION GAP WAS CONFIRMED BY INSERTING A SPACER BLOCK 10 MM (ACTUAL MEASUREMENT 19 MM), AND THE GAP WAS MEASURED IN EXTENSION AND FLEXION WITH THE LANTERN. THE EXTENSION GAP VALUE (CENTER VALUE 7MM) AND FLEXION GAP VALUE (CENTER VALUE 18MM) WERE INDICATED. THE OPERATION WAS COMPLETED WITH CONVENTIONAL INSTRUMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21021578 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K200892. One submission often covers many models, so check the product and lot details in each record.

  • Z-1630-2026 · Feb 10, 2026 · Open, Classified
    Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System
  • Z-1629-2026 · Feb 10, 2026 · Open, Classified
    Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System

Other Orthopedic Stereotaxic Instrument devices

All Orthopedic Stereotaxic Instrument devices →