HydroFrame 18 Advanced 7118-0931Recalls & reported problems
Microvention, Inc. · V-Trak · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00811425029770), received between Jan 21, 2017 and Jan 21, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Difficult to Remove” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Difficult to Remove1
- Separation Failure1
- Stretched1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 21, 2017Difficult to Remove; Stretched; Separation Failure; Difficult to Remove; Stretched; Separation Failure
IT WAS REPORTED THAT COIL EMBOLIZATION TREATMENT WAS BEING PERFORMED ON A SPLENIC ARTERY ANEURYSM. AFTER THE COIL WAS PLACED IN THE ANEURYSM, IT WOULD NOT DETACH. THE V-GRIP CONTROLLER WAS EXCHANGED, BUT THE COIL STILL WOULD NOT DETACH. AN ATTEMPT WAS MADE TO REMOVE THE ENTIRE COIL; HOWEVER, IT WAS LOCKED IN POSITION IN THE ANEURYSM. THE DELIVERY WIRE WAS TORQUED AT THE PROXIMAL END AND, BELIEVING THE COIL HAD DETACHED, THE DELIVERY WIRE WAS WITHDRAWN; HOWEVER, THE DISTAL PART OF THE DELIVERY WIRE STRETCHED AND THEN BROKE. A FEW CENTIMETERS OF THE HYDROFRAME PLUS A FEW CENTIMETERS OF THE DISTAL DELIVERY WIRE COIL REMAINED IN THE SPLENIC ARTERY. NO ADDITIONAL INTERVENTION WAS PERFORMED AND THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT. THE CURRENT PATIENT STATUS IS REPORTED TO BE FINE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6268846 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page