Skip to content

AMPLATZER™ 9-AVP2-004Recalls & reported problems

Aga Medical Corporation · Vascular Plug II · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00811806010359), received between Jun 11, 2018 and Dec 13, 2023 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1
  • Patient Device Interaction Problem1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT ON (B)(6) 2023, A 4MM AMPLATZER VASCULAR PLUG II WAS SELECTED FOR AN IMPLANT. DURING THE PROCEDURE, A NON-ABBOTT DELIVERY CATHETER WERE IN PLACE. THE PLUG WAS IMPLANTED, AND THE LEAK WAS OBSERVED 3 MINUTES LATER. THE CONTRAST AGENT STILL PASSES THROUGH THE TARGET VESSEL DUE TO POOR ADHERENCE. THE SIZE OF THE VESSEL PLUG IS NOT SUITABLE FOR THE SHAPE OF THE ARTERIAL DUCT. SO THE PHYSICIAN REPLACED WITH A NON-ABBOTT DEVICE TO COMPLETE THE PROCEDURE. THE PATIENT WAS STABLE, NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18322634 ↗
  • InjuryMar 6, 2019Migration; Migration

    RELATED MANUFACTURER NUMBER: 2135147-2019-00064. ON AN UNKNOWN DATE, AN 4 AMPLATZER VASCULAR PLUG II SELECTED FOR IMPLANT. THE DEVICE WAS DEPLOYED, BUT MIGRATED AFTER DEPLOYMENT. AN 6MM AMPLATZER VASCULAR PLUG II WAS THEN SELECTED BUT THAT DEVICE ALSO MIGRATED. AN 8MM AMPLATZER VASCULAR PLUG II WAS THEN SUCCESSFULLY IMPLANTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8397525 ↗
  • InjuryJun 11, 2018Unintended Movement; Unintended Movement

    IN (B)(6) 2018, 4 MM AMPLATZER VASCULAR PLUG II WAS IMPLANTED IN THE PATIENT DUCTUS ARTERIOUS. THE DAY AFTER IMPLANT, THE PATIENT PRESENTED WITH SYMPTOMS OF DEVICE EMBOLIZATION. ON (B)(6) 2018, PERCUTANEOUS RETRIEVAL OF THE DEVICE WAS PERFORMED AND REPLACED WITH A 6 MM AMPLATZER DUCT OCCLUDER II (LOT NUMBER: 6277108). THE PATIENT IS REPORTED TO BE STABLE. THE USER BELIEVES THEY UNDERSIZED THE DEVICE.

    Patient problems: No Information

    MAUDE report 7586952 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →