AMPLATZER™ 9-PLUG-010Recalls & reported problems

Aga Medical Corporation · Vascular Plug · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00811806011479), received between Mar 2, 2023 and Mar 3, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance2
  • Premature Separation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 3, 2025Difficult to Advance; Premature Separation; Difficult to Advance; Premature Separation

    IT WAS REPORTED THAT ON (B)(6) 2025, A 10MM AMPLATZER VASCULAR PLUG WAS CHOSEN FOR A ENDOVASCULAR AORTIC REPAIR (EVAR) CASE FOR INTERNAL ILIAC ARTERY EMBOLIZATION USING A 6F NON-ABBOTT DELIVERY SYSTEM. DURING PROCEDURE, THE DELIVERY SHEATH WAS ENGAGED IN THE CONTRALATERAL INTERNAL ILIAC ARTERY VIA THE LEFT FEMORAL APPROACH. WHILE DELIVERING THE DEVICE, THERE WAS SOME RESISTANCE DUE TO SEVERE TORTUOSITY. ALTHOUGH IT SEEMED POSSIBLE TO DELIVER, THE PUSHER WIRE WAS UNINTENTIONALLY ROTATED COUNTERCLOCKWISE DURING DELIVERY, CAUSING IT TO DETACH INSIDE THE CATHETER. THE CATHETER WAS REMOVED AND RETRIEVED. ANOTHER ATTEMPT TO MOUNT IT WAS UNSUCCESSFUL AND IT BECAME CONTAMINATED, A NEW 10MM AMPLATZER VASCULAR PLUG WAS CHOSEN AND COMPLETED THE PROCEDURE. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. THE PATIENT WAS REPORTED STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21506435 ↗
  • MalfunctionMar 2, 2023Difficult to Advance; Premature Separation; Difficult to Advance; Premature Separation

    IT WAS REPORTED THAT ON (B)(6) 2023, A 10MM AMPLATZER VASCULAR PLUG WAS SELECTED FOR IMPLANT USING A 6F NON-ABBOTT DELIVERY SYSTEM, WITH AN INNER DIAMETER OF 0.07 INCHES. IT WAS NOTED DURING PROCEDURE THAT THE PLUG WAS UNABLE TO BE ADVANCED. THE DECISION WAS MADE TO REMOVE THE PLUG FROM THE PATIENT'S ANATOMY. IT WAS NOTED AFTER SUCCESSFUL REMOVAL OF THE PLUG, THAT THE PLUG BECAME PREMATURELY DETACHED FROM THE DELIVERY CABLE. THE 10MM AMPLATZER VASCULAR PLUG HAD BEEN SIZED SELECTING A PLUG THAT IS 30 TO 50% LARGER THAN THE BLOOD VESSEL DIAMETER. THE PATIENT DOES NOT HAVE A TORTUOUS ANATOMY AND THE PLUG WAS NOT MISHANDLED OR DAMAGE AT ANY POINT DURING DEVICE PREPARATION. THE DECISION WAS MADE TO REPLACE THE 10MM AMPLATZER VASCULAR PLUG WITH AN UNKNOWN NON-ABBOTT DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16474422 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →