Skip to content

AMPLATZER™ 9-PLUG-012Recalls & reported problems

Aga Medical Corporation · Vascular Plug · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00811806011486), received between Nov 10, 2022 and Dec 4, 2023 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Migration or Expulsion of Device1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Aneurysm1
  • Angina1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2023, A 12MM AMPLATZER VASCULAR PLUG WAS IMPLANTED IN THE LEFT SUBCLAVIAN ARTERY (LSA) FOR THORACIC ENDOVASCULAR ANEURYSM REPAIR (TEVAR) WITH A 8F NON-ABBOTT GUIDING CATHETER. IT WAS REPORTED THE VESSEL DIAMETER WAS 8MM AND ECHOCARDIOGRAPHY CONFIRMED SUCCESSFUL IMPLANT WITH COMPLETE OCCLUSION OF BLOOD FLOW. AFTER 1 MONTH POST-PROCEDURE ON (B)(6) 2023, THE PATIENT EXPERIENCED PERSISTENT CHEST DISCOMFORT. IT WAS LATER REPORTED THAT ON (B)(6) 2023, AN ECHOCARDIOGRAPHY REVEALED AN ENDOLEAK OF BLOOD FLOW IN THE LSA. IT WAS THEN REPORTED ON (B)(6) 2023, A DECISION WAS MADE TO ADDITIONALLY IMPLANT A NON-ABBOTT DEVICE IN THE LSA. THE PATIENT STATUS WAS REPORTED AS STABLE.

    Patient problems: Aneurysm; Angina

    MAUDE report 18256836 ↗
  • InjuryAug 14, 2023Migration or Expulsion of Device; Migration or Expulsion of Device

    THE ARTICLE, ¿A CHALLENGING INTERVENTIONAL PROCEDURE: TRANSCATHETER CLOSURE OF TUBULAR PATENT DUCTUS ARTERIOSUS IN PATIENTS WITH PULMONARY HYPERTENSION¿, WAS REVIEWED. THE ARTICLE PRESENTS A CASE STUDY OF AN 1-YEAR-OLD, 6.6KG PATIENT WITH THE FOLLOWING PATENT DUCTUS ARTERIOSUS (PDA) MEASUREMENTS: MINIMAL DUCTUS DIAMETER = 7.9MM, AMPULLA DIAMETER = 7.8MM, DUCTAL LENGTH = 12.9MM. IT WAS REPORTED ON AN UNKNOWN DATE, A 12MM AMPLATZER VASCULAR PLUG I WAS CHOSEN FOR IMPLANT TO OCCLUDE A PDA. AFTER IMPLANT, IT WAS NOTED THE DEVICE HAD EMBOLIZED INTO THE PULMONARY ARTERY. A DECISION WAS MADE TO RETRIEVE AND EXPLANT THE EMBOLIZED DEVICE VIA TRANSCATHETER SNARE. A NEW 14MM AMPLATZER VASCULAR PLUG I WAS CHOSEN FOR IMPLANT. THE SECOND DEVICE ALSO EMBOLIZED INTO THE PULMONARY ARTERY AFTER RELEASE AND WAS ALSO EXPLANTED VIA TRANSCATHETER SNARE. A DECISION WAS MADE TO ADMIT THE PATIENT FOR SURGICAL LIGATION OF THE PDA. [THE PRIMARY AND CORRESPONDING AUTHOR IS ILKER KEMAL YUCEL, DEPARTMENT OF PEDIATRIC CARDIOLOGY, UNIVERSITY OF HEALTH SCIENCES DR. SIYAMI ERSEK THORACIC AND CARDIOVASCULAR SURGERY TRAINING AND RESEARCH HOSPITAL, ISTANBUL, TURKEY, [email protected]]. PERIPROCEDURAL COMPLICATI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17540973 ↗
  • MalfunctionNov 10, 2022Premature Separation; Premature Separation

    IT WAS REPORTED THAT ON (B)(6) 2022, A 12MM AMPLATZER VASCULAR PLUG WAS SELECTED FOR IMPLANTATION USING A 6F NON-ABBOTT SHEATH. DURING THE PROCEDURE, THE DEVICE WAS RE-SHEATHED SEVERAL TIMES. DURING THE RE-SHEATHING, THE DEVICE DETACHED INSIDE THE DELIVERY SYSTEM. THE DELIVERY SYSTEM WAS REMOVED FROM PATIENT ANATOMY. THE PHYSICIAN WAS UNSUCCESSFUL AT RE-ATTACHING THE DEVICE TO THE DELIVERY CABLE. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15775990 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →