AMPLATZER™ 9-PLUG-016Recalls & reported problems

Aga Medical Corporation · Vascular Plug · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00811806011509), received between Jul 1, 2022 and Dec 4, 2023 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration or Expulsion of Device2
  • Patient Device Interaction Problem1
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 4, 2023Migration or Expulsion of Device; Separation Problem; Migration or Expulsion of Device; Separation Problem

    IT WAS REPORTED THAT ON 10 NOVEMBER 2023, A 16MM AMPLATZER VASCULAR PLUG WAS SELECTED FOR AN IMPLANT FOR EMBOLIZATION OF SUBCLAVIAN ARTERY AS PART OF COMPLEX THORACIC ENDOVASCULAR AORTIC REPAIR (TEVAR) PROCEDURE. PLUG WAS UNSHEATHED TO SUBCLAVIAN ARTERY VIA BRACHIAL PUNCTURE AND 6F SHEATH. SIZING/POSITIONING CONFIRMED, PLUG WELL APPOSED TO VESSEL. PUSHER WIRE TURNED COUNTERCLOCKWISE MANY TIMES TO DETACH, (COULD NOT RECALL HOW MANY TIMES THE PUSHER WAS TURNED) BUT PLUG WOULD NOT DETACH. AFTER SOME MOVEMENT BACK AND FORTH WITH THE PUSHER WIRE, PLUG DETACHED. ON LATER IMAGING (6 WEEK FOLLOW UP SCAN) IT APPEARS THE PLUG HAS PARTIALLY PUSHED THROUGH SUBCLAVIAN ARTERY. PATIENT HAS BEEN DISCHARGED AND NO FURTHER COMMENTS FROM SURGEON AROUND THE DEVICE OR CONCERNS FOR THE PATIENT EXCEPT TO REPORT UNUSUAL FINDINGS ON POST IMAGING.

    Patient problems: Insufficient Information

    MAUDE report 18256695 ↗
  • MalfunctionOct 10, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    THE ARTICLE, ¿INTERMEDIATE AND LONG-TERM FOLLOW-UP OF TRANSCATHETER CLOSURE OF CONGENITAL CORONARY CAMERAL FISTULAS IN INFANTS AND CHILDREN: EXPERIENCE FROM A SINGLE CENTER¿, WAS REVIEWED.THE ARTICLE PRESENTED A CASE STUDY OF A 4-YEAR-OLD, 15KG PATIENT WITH A CORONARY CAMERAL FISTULA AND CORONARY ARTERY ANEURYSM. IT WAS REPORTED THAT ON AN UNKNOWN DATE, A 16MM AMPLATZER VASCULAR PLUG WAS CHOSEN FOR IMPLANT TO OCCLUDE A CORONARY CAMERAL FISTULA. IT WAS REPORTED 3 MONTHS POST-PROCEDURE, A LARGE RESIDUAL SHUNT WAS NOTED. THERE WAS NO TREATMENT REPORTED. THE ARTICLE CONCLUDED THAT TCC OF CCFS IN INFANTS AND CHILDREN APPEARS TO BE EFFECTIVE AND IS ASSOCIATED WITH A RELATIVELY LOW COMPLICATION RATE. LARGE CCFS AND GIANT CORONARY ARTERY ANEURYSMS (CAAS) REPRESENT A HIGHER RISK OF BOTH ACUTE AND INTERMEDIATE AND LONG-TERM ADVERSE EVENTS AFTER CLOSURE. [THE PRIMARY AND CORRESPONDING AUTHOR WAS ZHIWEI ZHANG, DEPARTMENT OF PEDIATRIC CARDIOLOGY, GUANGDONG CARDIOVASCULAR INSTITUTE, GUANGDONG PROVINCIAL PEOPLE¿S HOSPITAL (GUANGDONG ACADEMY OF MEDICAL SCIENCES), SOUTHERN MEDICAL UNIVERSITY, GUANGDONG PROVINCIAL KEY LABORATORY OF SOUTH CHINA STRUCTURAL HEART DISEASE, GUANGZHOU, 510100, CHINA, WITH

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17908661 ↗
  • InjuryJul 1, 2022Migration or Expulsion of Device; Migration or Expulsion of Device

    IT WAS REPORTED THAT A 16MM AMPLATZER VASCULAR PLUG WAS SELECTED FOR AN IMPLANT ON (B)(6) 2022. THE DEVICE WAS SUCCESSFULLY IMPLANTED. A SECOND UNKNOWN PLUG WAS PLACED PROXIMAL TO THE 16MM AMPLATZER VASCULAR PLUG. THE UNKNOWN PLUG BUMPED INTO THE 16MM AMPLATZER ,CAUSING IT TO MIGRATE FROM THE RIGHT HEPATIC VEIN AND ENDED UP IN THE RIGHT VENTRICLE. THE AMPLATZER DEVICE SUCCESSFULLY RETRIEVED VIA JUGULAR ACCESS & SNARE. THE UNKNOWN PLUG REMAINED IMPLANTED. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14891728 ↗

Other Device, Vascular, for Promoting Embolization devices