ELI 380 ELI380-DCS43Recalls & reported problems
Mortara Instrument, Inc. · Electrocardiograph · Class 2
All models of ELI 380 (11)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00812345026948), received between Apr 23, 2021 and Apr 23, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Wireless Communication Problem” (1 report).
1 FDA recall record name this device's identifier.
3 other recalls cover devices cleared under the same FDA submission (K142105); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Nov 11, 2020Class IITerminated
The radio within the device can become disassociated with the wireless access point.
Z-0553-2021 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Wireless Communication Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2021Wireless Communication Problem; Wireless Communication Problem
HILLROM RECEIVED A REPORT FROM THE ACCOUNT STATING THAT THE ELI 380 DEVICE WOULD NOT CONNECT TO THE WIRELESS NETWORK. THE DEVICE WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11716669 ↗
FDA clearance and classification
Cleared through Special 510(k) K142105 on Sep 5, 2014 (decision: Substantially Equivalent). Submitted by Mortara Instrument, Inc. as “Eli 380 Electrocardiograph”. The FDA review took 35 days. FDA record ↗
The FDA regulates product code DPS (electrocardiograph) under 21 CFR 870.2340 as a Class II device. 21 CFR 870.2340 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K142105. One submission often covers many models, so check the product and lot details in each record.
- Z-1013-2023 · Oct 31, 2022 · TerminatedBaxter Electrocardiograph, ELI 380 -DCS21
- Z-1483-2021 · Mar 31, 2021 · TerminatedELI 380 Electrocardiograph, with AM12M Acquisition Module accessory. Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediat
- Z-1796-2015 · May 29, 2015 · TerminatedELI 380 Electrocardiograph. The ELI 380 is intended to be a high-performance, multichannel resting electrocardiograph. As a resting electrocardiograph, the ELI 380 simultaneously a
Common questions about the ELI 380 ELI380-DCS43
Has the ELI 380 ELI380-DCS43 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0553-2021 (Class II), began on Nov 11, 2020: The radio within the device can become disassociated with the wireless access point. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the ELI 380 ELI380-DCS43?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 23, 2021. That report lists device problem “Wireless Communication Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ELI 380 ELI380-DCS43?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ELI 380 ELI380-DCS43 FDA approved?
It was cleared through Special 510(k) K142105 on Sep 5, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the ELI 380 ELI380-DCS43, and is it still sold?
The FDA's device database lists Mortara Instrument, Inc. as the company responsible for it (brand ELI 380, model ELI380-DCS43). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Electrocardiograph devices
- ELI 380 ELI380-DCS11Mortara Instrument, Inc. · Electrocardiograph1 recall1 death70 reports
- Cardiac Workstation Cardiac Workstation 7000Philips Medizin Systeme Böblingen GmbH · Electrocardiograph2 recalls16 reports
- Cardiac Workstation Cardiac Workstation 5000Philips Medizin Systeme Böblingen GmbH · Electrocardiograph2 recalls1 report
- Q-Stress QSERV-6AA-XXXXBMortara Instrument, Inc. · Electrocardiograph2 recalls1 report
- MAC VU360GE Medical Systems Information Technologies, Inc. · Electrocardiograph2 recalls1 report
- XScribe Cardiopulmonary Kit 41000-030-62Mortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Q-Stress QS6-AXTDSMortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Welch Allyn, Inc. 901007Welch Allyn, Inc. · Electrocardiograph2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page