ELI 380 ELI380-DCX21Recalls & reported problems
Mortara Instrument, Inc. · Electrocardiograph · Class 2
All models of ELI 380 (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00812345026986), received between Sep 29, 2020 and Apr 5, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Wireless Communication Problem” (5 reports).
1 FDA recall record name this device's identifier.
3 other recalls cover devices cleared under the same FDA submission (K142105); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Nov 11, 2020Class IITerminated
The radio within the device can become disassociated with the wireless access point.
Z-0553-2021 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Wireless Communication Problem5
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Patient Involvement3
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 5, 2024Wireless Communication Problem; Wireless Communication Problem
CUSTOMER REPORTED DEVICE UNABLE TO BE PINGED TO THE NETWORK AND CONNECTION IS DROPPING. HERE WAS NO PATIENT/USER INJURY REPORTED. THIS EVENT WAS CAPTURED UNDER HILLROM COMPLAINT REF (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19048931 ↗MalfunctionApr 4, 2024Wireless Communication Problem; Wireless Communication Problem
CUSTOMER REPORTED DEVICE UNABLE TO BE PINGED TO THE NETWORK AND CONNECTION IS DROPPING. HERE WAS NO PATIENT/USER INJURY REPORTED. THIS EVENT WAS CAPTURED UNDER HILLROM COMPLAINT REF (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19046852 ↗MalfunctionMay 4, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE ELI380 FAILED TO CONNECT TO THE NETWORK. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14276846 ↗MalfunctionOct 14, 2020Wireless Communication Problem; Wireless Communication Problem
HILLROM RECEIVED A REPORT FROM THE ACCOUNT STATING THAT THE ELI 380 DEVICE WOULD DROP CONNECTION TO THE WIRELESS NETWORK. THE DEVICE WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10679121 ↗MalfunctionOct 14, 2020Wireless Communication Problem; Wireless Communication Problem
HILLROM RECEIVED A REPORT FROM THE ACCOUNT STATING THAT THE ELI 380 DEVICE WOULD DROP CONNECTION TO THE WIRELESS NETWORK. THE DEVICE WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Patient Involvement
MAUDE report 10678988 ↗MalfunctionSep 29, 2020Wireless Communication Problem; Wireless Communication Problem
HILLROM RECEIVED A REPORT FROM THE ACCOUNT STATING THAT THE ELI 380 DEVICE WAS HAVING DIFFICULTY REMAINING CONNECTED TO THE WIRELESS NETWORK. THE DEVICE WAS LOCATED IN THE ACCOUNT AT THE TIME OF THE EVENT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT#: (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10602847 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K142105. One submission often covers many models, so check the product and lot details in each record.
- Z-1013-2023 · Oct 31, 2022 · TerminatedBaxter Electrocardiograph, ELI 380 -DCS21
- Z-1483-2021 · Mar 31, 2021 · TerminatedELI 380 Electrocardiograph, with AM12M Acquisition Module accessory. Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediat
- Z-1796-2015 · May 29, 2015 · TerminatedELI 380 Electrocardiograph. The ELI 380 is intended to be a high-performance, multichannel resting electrocardiograph. As a resting electrocardiograph, the ELI 380 simultaneously a
Other Electrocardiograph devices
- ELI 380 ELI380-DCS11Mortara Instrument, Inc. · Electrocardiograph1 recall1 death70 reports
- Cardiac Workstation Cardiac Workstation 7000Philips Medizin Systeme Böblingen GmbH · Electrocardiograph2 recalls17 reports
- Q-Stress QSERV-6AA-XXXXBMortara Instrument, Inc. · Electrocardiograph2 recalls1 report
- Cardiac Workstation Cardiac Workstation 5000Philips Medizin Systeme Böblingen GmbH · Electrocardiograph2 recalls1 report
- MAC VU360GE Medical Systems Information Technologies, Inc. · Electrocardiograph2 recalls1 report
- XScribe Cardiopulmonary Kit 41000-030-60Mortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Welch Allyn, Inc. 901049Welch Allyn, Inc. · ELECTROCARDIOGRAPH2 recalls0 reports
- Welch Allyn, Inc. 901133Welch Allyn, Inc. · Electrocardiograph2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page