Mortara Instrument, Inc. · Cardiac Stress Testing System · Class 2

All models of Q-Stress (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00812345027563), received between Oct 31, 2023 and Jul 8, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Patient Data Problem” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K001492, K082173, K152944); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 11, 2024Class IIOpen, Classified

    There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.

    Z-0321-2025 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Patient Data Problem4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 8, 2026Patient Data Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25756920 ↗
  • MalfunctionJun 8, 2026Patient Data Problem

    BAXTER RECEIVED A REPORT STATING THAT AN EXAM HAD INCORRECT PATIENT DATA IN Q-STRESS. THE EVENT OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25437196 ↗
  • MalfunctionAug 6, 2024Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTED THAT STUDY PATIENT INFORMATION CHANGED AT THE BEGINNING OF THE STUDY. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT HAS BEEN CAPTURED UNDER HILLROM COMPLAINT #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19921347 ↗
  • MalfunctionOct 31, 2023Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTED THE QSTRESS 6 HAD A NUMBER OF ISSUES WITH TEST REPORTS, INCLUDING THE PATIENT NAME SWITCHED INSIDE THE EXAM WHEN USED IN OBSERVATION MODE, AND WAS UNABLE TO USE SECONDARY PATIENT ID WHEN USING MWL ORDER. THERE WAS NO INJURY REPORTED. THIS EVENT HAS BEEN CAPTURED UNDER HILLROM COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18038835 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K001492, K082173, K152944. One submission often covers many models, so check the product and lot details in each record.

  • Z-0325-2025 · Oct 11, 2024 · Open, Classified
    Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer
  • Z-0706-2024 · Dec 8, 2023 · Open, Classified
    XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-BXXXX; b) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-DXXXX; c) XS6
  • Z-0705-2024 · Dec 8, 2023 · Open, Classified
    Q-Stress Cardiac Stress Testing System, version 6, with below product name / product codes: a) QS6 AM12Q KIT AHA XML / QS6-BXXCS: b) QS6 AM12Q KIT BCRT IEC / QS-6CC-DGCXX: c) QS6
  • Z-1817-2020 · Mar 18, 2020 · Terminated
    Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wire
  • Z-1757-2010 · Mar 30, 2009 · Terminated
    Treadmill models TM55, TM65, and CR60 (brand names include Quinton, Burdick, and Cardiac Science).

Other Electrocardiograph devices

All Electrocardiograph devices →