Skip to content

Q-Stress QS6-MLTC1Recalls & reported problems

Mortara Instrument, Inc. · Cardiac Stress Testing System · Class 2

All models of Q-Stress (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00812345027631), received between Sep 17, 2025 and Sep 17, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Patient Data Problem” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K001492, K082173, K152944); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Data Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2025Patient Data Problem; Patient Data Problem

    THE CUSTOMER CONTACTED BAXTER TO REPORT THE NURSE HAD SAVED AN EXAM WITH THE INCORRECT PATIENT INFORMATION. THERE WAS NO ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23070644 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K001492, K082173, K152944. One submission often covers many models, so check the product and lot details in each record.

  • Z-0321-2025 · Oct 11, 2024 · Open, Classified
    Hillrom Welch Allyn Q-Stress Cardiac Stress Testing System; cardiac stress test
  • Z-0325-2025 · Oct 11, 2024 · Open, Classified
    Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer
  • Z-0706-2024 · Dec 8, 2023 · Open, Classified
    XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-BXXXX; b) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-DXXXX; c) XS6
  • Z-0705-2024 · Dec 8, 2023 · Open, Classified
    Q-Stress Cardiac Stress Testing System, version 6, with below product name / product codes: a) QS6 AM12Q KIT AHA XML / QS6-BXXCS: b) QS6 AM12Q KIT BCRT IEC / QS-6CC-DGCXX: c) QS6
  • Z-1817-2020 · Mar 18, 2020 · Terminated
    Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wire
  • Z-1757-2010 · Mar 30, 2009 · Terminated
    Treadmill models TM55, TM65, and CR60 (brand names include Quinton, Burdick, and Cardiac Science).

Other Electrocardiograph devices

All Electrocardiograph devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page