Skip to content

Q-Stress QS6-MTTC1Recalls & reported problems

Mortara Instrument, Inc. · Cardiac Stress Testing System · Class 2

All models of Q-Stress (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00812345027655), received between Sep 20, 2023 and Jan 23, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K001492, K082173, K152944); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loose or Intermittent Connection1
  • Patient Data Problem1

Patient problems

  • Abrasion1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2024Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTED THE INCORRECT PATIENT KEEPS APPEARING ON COMPLETED EXAMS. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT HAS BEEN CAPTURED UNDER HILLROM COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18562761 ↗
  • MalfunctionSep 20, 2023Loose or Intermittent Connection; Loose or Intermittent Connection

    THE CUSTOMER REPORTED THAT DURING THE USE OF THE QSTRESS DEVICE IN CONJUNCTION WITH A TREADMILL, AFTER THE DEVICE PROGRESSED THE TEST/TREADMILL FROM STAGE 1 TO STAGE 2, THE PATIENT REQUESTED THE CAREGIVER TO STOP THE TREADMILL. OF NOTE, THIS PATIENT WAS REPORTED TO HAVE A "PHOBIA" OF TREADMILLS DUE TO THE PATIENT FALLING FROM A TREADMILL AT A PREVIOUS UNDISCLOSED TIME. WHEN THE CAREGIVER ATTEMPTED TO STOP THE TEST/TREADMILL VIA THE QSTRESS STOP BUTTON, THERE WAS NO RESPONSE, THE PATIENT THEN FELL SUSTAINING A MINOR ABRASION ON HER LEFT HIP. UPON THE ARRIVAL OF A SECOND CAREGIVER, THE SECOND CAREGIVER PRESSED THE ESTOP BUTTON ON THE TREADMILL TO STOP THE TREADMILL. THE PATIENT WAS ASSESSED AT THE TIME OF THE EVENT AS WELL AS AT A LATER TIME BY THE PATIENT'S PERSONAL HEALTHCARE PROVIDER. NO MEDICAL INTERVENTION WAS REQUIRED AT EITHER ASSESSMENT. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: Abrasion

    MAUDE report 17783294 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K001492, K082173, K152944. One submission often covers many models, so check the product and lot details in each record.

  • Z-0321-2025 · Oct 11, 2024 · Open, Classified
    Hillrom Welch Allyn Q-Stress Cardiac Stress Testing System; cardiac stress test
  • Z-0325-2025 · Oct 11, 2024 · Open, Classified
    Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer
  • Z-0705-2024 · Dec 8, 2023 · Open, Classified
    Q-Stress Cardiac Stress Testing System, version 6, with below product name / product codes: a) QS6 AM12Q KIT AHA XML / QS6-BXXCS: b) QS6 AM12Q KIT BCRT IEC / QS-6CC-DGCXX: c) QS6
  • Z-0706-2024 · Dec 8, 2023 · Open, Classified
    XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-BXXXX; b) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-DXXXX; c) XS6
  • Z-1817-2020 · Mar 18, 2020 · Terminated
    Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wire
  • Z-1757-2010 · Mar 30, 2009 · Terminated
    Treadmill models TM55, TM65, and CR60 (brand names include Quinton, Burdick, and Cardiac Science).

Other Electrocardiograph devices

All Electrocardiograph devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page