Skip to content

CORMIX MS64050Recalls & reported problems

Metrix Company, The · EMPTY EVA CONTAINER With 2 Ports · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

16 other recalls cover devices cleared under the same FDA submission (K030888); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Oct 9, 2024Open, Classified

    A limited number of IV bags have been found to leak during filling.

    Z-0334-2025 · Root cause: Under Investigation by firm · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K030888. One submission often covers many models, so check the product and lot details in each record.

  • Z-0336-2025 · Oct 9, 2024 · Open, Classified
    SECURE 100 mL EMPTY EVA CONTAINER With 2 Ports (Bulk Packed), REF 66041; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0344-2025 · Oct 9, 2024 · Open, Classified
    Imed Products Imed EVA BAG, 500 mL, 2-PORTS, REF IM68050; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0341-2025 · Oct 9, 2024 · Open, Classified
    SECURE 1000 mL EMPTY EVA CONTAINER With 2 Ports, REF 66055; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0339-2025 · Oct 9, 2024 · Open, Classified
    SECURE 250 mL EMPTY EVA CONTAINER With 2 Ports, REF 66045; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0342-2025 · Oct 9, 2024 · Open, Classified
    Imed Products Imed EVA BAG, 150 mL, 2-PORTS, REF IM38040; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0330-2025 · Oct 9, 2024 · Open, Classified
    VYGON 100 mL EMPTY EVA CONTAINER With 2 Ports, REF CMS-9601; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0338-2025 · Oct 9, 2024 · Open, Classified
    SECURE 50 mL EMPTY EVA CONTAINER With 2 Ports (Bulk Packed) , REF 66043; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0337-2025 · Oct 9, 2024 · Open, Classified
    SECURE 50 mL EMPTY EVA CONTAINER With 2 Ports, REF 66042; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0340-2025 · Oct 9, 2024 · Open, Classified
    SECURE 500 mL EMPTY EVA CONTAINER With 2 Ports, REF 66050; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0331-2025 · Oct 9, 2024 · Open, Classified
    VYGON 250 mL EMPTY EVA CONTAINER With 2 Ports, REF CMS-9602; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0343-2025 · Oct 9, 2024 · Open, Classified
    Imed Products Imed EVA BAG, 250 mL, 2-PORTS, REF IM68045; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0333-2025 · Oct 9, 2024 · Open, Classified
    CORMIX 250 mL EMPTY EVA CONTAINER With 2 Ports, REF MS64045; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0332-2025 · Oct 9, 2024 · Open, Classified
    CORMIX 150 mL EMPTY EVA CONTAINER With 2 Ports, REF MS64040; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0345-2025 · Oct 9, 2024 · Open, Classified
    Imed Products Imed EVA BAG, 1000 mL, 2-PORTS, REF IM68055; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
  • Z-0335-2025 · Oct 9, 2024 · Open, Classified
    SECURE 100 mL EMPTY EVA CONTAINER With 2 Ports, REF 66040; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.

…and 1 more.

Other Container, I.V. devices

All Container, I.V. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page