SECURE® 66050Recalls & reported problems
Metrix Company, The · EMPTY EVA CONTAINER With 2 Ports · Class 2
All models of SECURE (16)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00812496011596), received between Jan 9, 2025 and Jan 9, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
16 other recalls cover devices cleared under the same FDA submission (K030888); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Oct 9, 2024Open, Classified
A limited number of IV bags have been found to leak during filling.
Z-0340-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 9, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER INITIALLY STATED, "OUR COMPOUNDING PHARMACIST LET US KNOW THAT THE 500ML EVA SECURE BAGS (REF 66050) ARE CONSISTENTLY LEAKING. THESE ONES WERE NOT PART OF THE RECALL FROM OCTOBER (THEY HAVE A "10" BETWEEN THE TWO PORTS)". UPON REQUEST FOR ADDITIONAL INFORMATION, THE CUSTOMER FURTHER REPORTED, "THE BAG WAS LEAKING FAIRLY VIGOROUSLY FROM THE BOTTOM OF THE BAG, ON THE OUTSIDE OF THE PORTS (SO, NOT IN BETWEEN THE PORTS). SHE THINKS, BUT IS NOT CERTAIN, THAT IT WAS THE SIDE OF THE BAG WITH THE ADDITION PORT. SHE WAS SPECIFIC THAT IT WAS A NUMBER 8 BAG." THIS LOT NUMBER (66050-A8460) WAS INVOLVED IN THE RECALL CORRECTION OF 2 PORT BAGS MANUFACTURED FROM 12/8/2023 TO 9/26/2024. PER RECALL PROCEEDINGS, THE FINISHED PRODUCT WAS SORTED AND REPACKAGED TO REMOVE CAVITY #6 FROM THE LOT. THE PRODUCT LEAK WAS IDENTIFIED BY THE FILLING PHARMACY, THUS NO PATIENT CONTACT OR ADVERSE EVENT WAS REPORTED. NO SAMPLES OR IMAGES OF THE LEAKING PRODUCT WERE AVAILABLE TO REVIEW. 43 UNUSED SAMPLES WERE RETURNED FROM LOT 66050-A8460 FOR METRIX INVESTIGATION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21112098 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K030888. One submission often covers many models, so check the product and lot details in each record.
- Z-0345-2025 · Oct 9, 2024 · Open, ClassifiedImed Products Imed EVA BAG, 1000 mL, 2-PORTS, REF IM68055; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0333-2025 · Oct 9, 2024 · Open, ClassifiedCORMIX 250 mL EMPTY EVA CONTAINER With 2 Ports, REF MS64045; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0330-2025 · Oct 9, 2024 · Open, ClassifiedVYGON 100 mL EMPTY EVA CONTAINER With 2 Ports, REF CMS-9601; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0339-2025 · Oct 9, 2024 · Open, ClassifiedSECURE 250 mL EMPTY EVA CONTAINER With 2 Ports, REF 66045; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0336-2025 · Oct 9, 2024 · Open, ClassifiedSECURE 100 mL EMPTY EVA CONTAINER With 2 Ports (Bulk Packed), REF 66041; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0331-2025 · Oct 9, 2024 · Open, ClassifiedVYGON 250 mL EMPTY EVA CONTAINER With 2 Ports, REF CMS-9602; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0343-2025 · Oct 9, 2024 · Open, ClassifiedImed Products Imed EVA BAG, 250 mL, 2-PORTS, REF IM68045; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0341-2025 · Oct 9, 2024 · Open, ClassifiedSECURE 1000 mL EMPTY EVA CONTAINER With 2 Ports, REF 66055; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0342-2025 · Oct 9, 2024 · Open, ClassifiedImed Products Imed EVA BAG, 150 mL, 2-PORTS, REF IM38040; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0337-2025 · Oct 9, 2024 · Open, ClassifiedSECURE 50 mL EMPTY EVA CONTAINER With 2 Ports, REF 66042; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0338-2025 · Oct 9, 2024 · Open, ClassifiedSECURE 50 mL EMPTY EVA CONTAINER With 2 Ports (Bulk Packed) , REF 66043; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0334-2025 · Oct 9, 2024 · Open, ClassifiedCORMIX 500 mL EMPTY EVA CONTAINER With 2 Ports, REF MS64050; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0332-2025 · Oct 9, 2024 · Open, ClassifiedCORMIX 150 mL EMPTY EVA CONTAINER With 2 Ports, REF MS64040; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0344-2025 · Oct 9, 2024 · Open, ClassifiedImed Products Imed EVA BAG, 500 mL, 2-PORTS, REF IM68050; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
- Z-0335-2025 · Oct 9, 2024 · Open, ClassifiedSECURE 100 mL EMPTY EVA CONTAINER With 2 Ports, REF 66040; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
…and 1 more.
Other Container, I.V. devices
- INTRAVIA 2B8011Baxter International Inc. · CONTAINER, I.V.127 reports
- INTRAVIA 2B8012Baxter International Inc. · CONTAINER, I.V.119 reports
- APEX™ 2112531B. Braun Medical Inc. · CONTAINER, I.V.69 reports
- APEX™ 2112530B. Braun Medical Inc. · CONTAINER, I.V.66 reports
- INTRAVIA 2B8013Baxter International Inc. · CONTAINER, I.V.64 reports
- SECURE® 66040Metrix Company, The · Container, I.V.1 recall8 reports
- MICROTEK 1Microtek Medical LLC · CONTAINER, I.V.1 recall8 reports
- SECURE® 66045Metrix Company, The · Container, I.V.1 recall7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.